10
EC170-EU-EN V1.1 7/13
Factory
Default
Settings
Type
Parameter
Default
Notes
P1.0
Select
°C/°F
°C
Temperature
units
P3.0
Revert
to
Factory
Default
settings
NO
Select
YES
to
revert
to
default
settings
P4.1
‐‐‐‐‐
Calibration
Data
for
Range
1
P4.2
Review
previous
Calibration
data
‐‐‐‐‐
Calibration
Data
for
Range
2
Specifications
Measurement
ranges
Range
1:
0.00
to
10.00
ppt
Salinity
Range
2:
10.1
to
70.0
ppt
Salinity
ppt
=
parts
per
thousand
1ppt
=
1
g/L
Salinity
Accuracy
Range
1:
1%
Full
Scale
±1digit
or
2%
Full
Scale
Range
2:
±1digit
Salinity
Resolution
Range
1:
0.01ppt
Range
2:
0.1ppt
Temperature
Accuracy
±0.5°C
(0.9°F)
Temperature
Resolution
0.1°C/°F
TDS
Factor
Built
‐
in
NaCl
conductivity
to
TDS
conversion
factor
Calibration
One
point
calibration
per
range
Auto
Power
OFF
After
20
minutes
of
inactivity
Data
Hold
Freezes
displayed
reading
Automatic
Temperature
Compensation
(ATC):
0
to
50°C
(32
to
122°F)
Waterproof
IP65
rated
Temperature
Coefficient
Built
‐
in
NaCl
temperature
coefficient
Operating
Conditions
Temperature:
0
to
50°C
(32
to
122°F);
<80%RH
Storage
Conditions
Temperature:
0
to
60°C
(32
to
140°F);
<90%RH
Normalization
Temperature
Fixed
at
25°C
(77°F)
Basic
status
indicators
Out
‐
of
‐
range
(
‐‐‐‐
)
and
low
battery
Power
Supply
Four
(4)
LR44
‘button’
batteries
Dimensions
Meter:
165
x
35
x
32mm
(6.5
x
1.4
x
1.3”)
LCD:
30
x
18mm
(1.2
x
0.7”)
Weight
115g
(4
oz.)
Copyright
©
2013
FLIR
Systems,
Inc.
All
rights
reserved
including
the
right
of
reproduction
in
whole
or
in
part
in
any
form
www.extech.com
www.
GlobalTestSupply
.com
Find Quality Products Online at: