Extech Instruments EA30 User Manual Download Page 5

                                                                                                                                     EA30  V2.0    06/06 

Data Hold 

Press the 

HOLD

 button to freeze the displayed reading. The 'H' hold icon will appear on 

the upper right-hand side of the display. Press the 

HOLD

 button again to return to normal 

operation (the 'H' hold icon will disappear). When hold function is enabled the analog bar 
graph will continue to display level changes. 

Peak Hold 

The Peak Hold feature allows the meter to captures light level changes that occur in as 
little time as 100us. Press the 

PEAK

 button momentarily and the ‘P+’ icon will appear on 

the LCD. The meter can now measure a fast acting light peak and display it on the LCD. 
To exit the Peak Hold mode, press and hold the 

PEAK

 button until the ‘P+’ icon 

disappears.  
NOTE: If the 

PEAK

 button is pressed 

momentarily

 rather then pressed and held when 

exiting the Peak mode, the meter will display ‘P-‘ which is a function reserved for future use 
and should be ignored. 

Relative mode 

Measurements can be displayed as a difference between actual light level and a stored 
reference value. To store a reading as a reference, press the 

REL

 button when the desired 

reference measurement is on the LCD (the triangle icon will appear steady in the upper 
left-hand corner of the LCD). All subsequent displayed readings will be “relative” to the 
stored reference value. For example, if the reference value is 100 and the actual light level 
is 125, the meter will display 25. To view the reference value, press the 

REL

 button again 

so that the triangle icon begins to flash. The displayed value will be the reference value. To 
exit the Relative mode, press and hold the 

REL

 button until the triangle icon disappears. 

Summary of Contents for EA30

Page 1: ...User s Guide EasyView Digital Light Meter Model EA30 ...

Page 2: ... prevent damage in transit This warranty does not apply to defects resulting from action of the user such as misuse improper wiring operation outside of specification improper maintenance or repair or unauthorized modification Extech specifically disclaims any implied warranties or merchantability or fitness for a specific purpose and will not be liable for any direct indirect incidental or conseq...

Page 3: ...ral accuracy CIE V function f1 ï 6 Measurement Repeatability 2 Temperature coefficient 0 1 per o C Sampling rate 13 3 times per second bargraph 1 3 times per second digital display Photo detector Silicon photo diode with spectral response filter Peak hold Capture light peaks to 100uS Operating conditions Temperature 32 to 104 o F 0 to 40 o C Humidity 80 RH Storage conditions Temperature 14 to 140 ...

Page 4: ...tored For overhead lighting the sensor can be placed on a desk or table top 2 The light being measured must encompass the entire white sensor dome 3 Read the light level on the LCD If the light measurement is out of range an OL indication will appear on the display If the OL appears select another range by pressing the RANGE button Selecting Lux or Foot candle units of measure Press the LUX button...

Page 5: ...EAK button is pressed momentarily rather then pressed and held when exiting the Peak mode the meter will display P which is a function reserved for future use and should be ignored Relative mode Measurements can be displayed as a difference between actual light level and a stored reference value To store a reading as a reference press the REL button when the desired reference measurement is on the...

Page 6: ...his manual Battery Replacement When the battery power falls low the BT symbol will appear on the lower left hand side of the LCD Replace the six 6 1 5 AAA batteries by removing the rear center battery compartment screw and accessing the battery compartment Observe polarity when placing the batteries in the compartment Ensure that the compartment cover is securely fastened when finished Calibration...

Page 7: ...or Stairs 150 300 15 30 Entrance Wash Room 200 750 20 75 Conference Reception Room 300 750 30 75 Cooking Room Dining Table 750 1 500 75 150 Clerical Work 750 1 500 75 150 Show Window 1 500 2 000 150 2000 Typing Drafting Store Hospital 75 150 7 15 Indoors 30 75 3 7 Emergency Stairs 150 200 15 20 Corridor Stairs 75 100 7 10 Stairs 200 300 20 30 Reception 100 150 10 15 Sick Room Warehouse 300 500 30 ...

Reviews: