background image

EXTECH

 

INSTRUMENTS

 

7

                                               

CO50

en

GB_V1.2

   

12/14

 

MANUALLY

 

STORE/RECALL

 

99

 

CO

 

READINGS

 

The

 

CO50

 

can

 

manually

 

store

 

a

 

CO

 

reading

 

with

 

a

 

press

 

of

 

the

 

SEL/RECALL

 

button.

 

The

 

LCD

 

shows

 

the

 

current

 

record

 

number

 

(01

99)

 

in

 

the

 

humidity

 

reading

 

area

 

each

 

time

 

a

 

reading

 

is

 

manually

 

stored

 

(a

 

short

 

beep

 

will

 

sound

 

each

 

time

 

a

 

reading

 

is

 

stored

 

also).

 

To

 

recall

 

stored

 

readings,

 

press

 

and

 

hold

 

the

 

SEL/RECALL

 

Button,

 

the

 

LCD

 

shows

 

the

 

number

 

of

 

the

 

record

 

with

 

its

 

CO

 

reading,

 

to

 

scroll

 

other

 

stored

 

readings

 

use

 

the

 

 

and

 

 

buttons.

 

To

 

clear

 

stored

 

data:

  

Access

 

the

 

Recall

 

mode

 

and

 

press

 

and

 

hold

 

the

 

SET

 

button

 

until

 

“Clr

 

“appears

 

on

 

LCD.

 

The

 

LCD

 

automatically

 

returns

 

to

 

normal

 

mode.

 

The

 

illustration

 

below

 

on

 

the

 

left

 

shows

 

a

 

manually

 

recorded

 

reading

 

(reading

 

number

 

‘05’

 

in

 

this

 

case);

 

the

 

middle

 

illustration

 

shows

 

the

 

recalled

 

reading;

 

the

 

illustration

 

on

 

the

 

right

 

shows

 

readings

 

cleared.

 

 

 

 

 

 

 

 

 

 

 

 

CO

 

CALIBRATION

 

It

 

is

 

recommended

 

that

 

the

 

meter

 

be

 

calibrated

 

monthly

 

in

 

a

 

clean

 

air

 

environment.

 

From

 

the

 

normal

 

measurement

 

mode,

 

press

 

and

 

hold

 

the

 

MODE

 

button

 

for

 

at

 

least

 

6

 

seconds

 

to

 

enter

 

calibration

 

mode.

 

The

 

LCD

 

shows

 

CAL

 

and

 

a

 

number

 

representing

 

the

 

time

 

remaining

 

in

 

the

 

calibration

 

(15

 

seconds).

 

The

 

meter

 

returns

 

to

 

the

 

normal

 

measurement

 

mode

 

automatically

 

when

 

calibration

 

is

 

complete.

 

Warning:

 

Ensure

 

that

 

no

 

CO

 

concentration

 

is

 

present

 

while

 

calibrating.

 

 

 

 

Reading

 

stored

 

Reading

 

Recalled

Reading

 

Cleared

 

Summary of Contents for CO50

Page 1: ...USER GUIDE Desktop Carbon Monoxide Monitor Model CO50...

Page 2: ...art measurement technology Easy to use intended for indoor air quality monitoring Bright multicolor LED representing Recording Green Alarm Red Startup Orange 99 data point memory User adjustable CO Al...

Page 3: ...DE Button with down arrow Decrease a value Review Maximum MAX reading Meter Description 1 Top LCD display for CO concentration 2 Middle LCD display for RH and temperature 3 Bottom LCD display Time and...

Page 4: ...er tone will sound The LCDs show the current CO value in ppm top ambient temperature relative humidity middle and the date and time bottom Toggle the temperature units by pressing and holding the C F...

Page 5: ...ss SET again to skip the Time setting Set the Time if desired per earlier instructions 4 A flashing digit will appear Use the up and down arrows to increase decrease the alarm threshold Press SEL RECA...

Page 6: ...m temporarily From the normal measurement mode press the SET MODE button at the same time the bell icon disappears After 4 minutes the alarm will turn on again automatically To comply with UL regulati...

Page 7: ...ntil Clr appears on LCD The LCD automatically returns to normal mode The illustration below on the left shows a manually recorded reading reading number 05 in this case the middle illustration shows t...

Page 8: ...distance Temperature range 10 to 60 o C 14 to 140 o F Temperature resolution 0 1 o C o F Temperature Accuracy 0 6 o C 1 2 o F RH range 10 to 90 RH RH Resolution 0 1 RH RH Accuracy 5 RH at 25 o C other...

Reviews: