2                                     433201       Ver. 1.04    11/99

 

 

3. FRONT PANEL DESCRIPTION  

 

1.  LED display 

2.  TYPE button 

3. 

o

C/

o

F button 

4.  COURSE output adjust 

5.  Battery compartment 

6.  Output/Input mini connector 

7.  Measure/Calibrate switch 

8.  FINE output adjust 

9.  AC Adaptor connector 

10.  On/Off switch 

 

 

 

 

 

4.  SETUP 

4.1  For continuous use, connect the AC/DC Adaptor/Recharger to a power outlet and to 

the “ADAPTOR” socket on the side of the meter. 

4.2  For portable operation using the 8.4V NiCad battery use the following procedure. 
4.3  Insert the NiCad battery into the meter housing by closing the case and inserting a 

coin or flat screwdriver into the slot at the top of the meter. Snap the cover in the 

direction of the arrow and then open the cover of the meter. Remove the battery cover 

and connect the snap terminals to the battery terminals. Close the battery cover 

firmly. 

4.4  Charge the battery for 24 to 48 hours before its first use. An overnight charge will 

enable continuous use for approximately 45 minutes. A full charge will provide one 

hour of continuous use. Turning the meter OFF when not in use will prolong battery 

life. 
 

5.  MEASURING TEMPERATURE 

5.1  Press the red power switch to the ON position. 
5.2  Select the “MEASURE” mode on the side panel slide switch 
5.3  Select the thermocouple type of your measuring probe via the faceplate “TYPE” 

button. An LED will light above the T/C type selected in the display. Each time the 

“TYPE” button is pressed the indicator will advance one type. 

5.4  Select 

o

C or 

o

F via the “C/F” button. 

5.5  Insert the thermocouple probe into the sub-miniature female connector on the side 

panel. Be sure to check that it is inserted with the proper polarity and make sure the 

proper type is selected in the display. 

5.6  Read the temperature on the LED display. 
 

6.  CALIBRATING THERMOCOUPLE DEVICES 

6.1  Press the red power SWITCH TO THE “on” POSITION. 
6.2  Select the “CALIBRATE” mode on the side panel slide switch. 
6.3  Select the thermocouple type of the device to be calibrated via the faceplate “TYPE” 

button. An LED will light above the T/C type selected in the display. Each time the 

“TYPE” button is pressed the indicator will advance one type. 

6.4  Select 

O

C or 

O

F via the “C/F” button. 

Summary of Contents for 433201

Page 1: ...ing angle and the meter automatically shuts off when the cover is closed A precision millivolt signal is either input from a thermocouple Measure mode or output to another thermocouple device Calibrate mode and the corresponding temperature according to standard temperature tables is displayed on the meter s 4 digit LED display 2 SPECIFICATIONS Calibrate Measure Resolution Accuracy Type J 58 to 18...

Page 2: ...arge will provide one hour of continuous use Turning the meter OFF when not in use will prolong battery life 5 MEASURING TEMPERATURE 5 1 Press the red power switch to the ON position 5 2 Select the MEASURE mode on the side panel slide switch 5 3 Select the thermocouple type of your measuring probe via the faceplate TYPE button An LED will light above the T C type selected in the display Each time ...

Page 3: ...ine battery may be used in place of the rechargeable 8 4V NiCad battery that comes with the unit 8 3 A flashing F or C on the LED display indicates a low battery condition 9 CALIBRATION REPAIR SERVICES Extech offers complete repair and calibration services for all of the products we sell For periodic calibration NIST certification or repair of any Extech product call customer service for details o...

Page 4: ...gulo de visión y el medidor se apaga automáticamente cuando la cubierta se cierra Se introduce una señal de milivatios precisa del termopar Modo de medición o sale a otro dispositivo termopar Modo calibración y la temperatura correspondiente de conformidad con tablas estándar de temperatura se muestra in la pantalla LED de 4 dígitos del medidor 2 ESPECIFICACIONES Calibración y Medición Resolución ...

Page 5: ... y conecte las terminales de broche en las terminales de la batería Cierre firmemente la cubierta de la batería 4 8 Cargue la batería NiCad durante 24 a 48 horas antes de su primer uso Una carga de una noche proporcionará energía para 45 minutos de operación continua aproximadamente Una carga completa proporcionará una hora de uso continuo Apagar el medidor cuando no esté en uso prolongará la vida...

Page 6: ... al dispositivo a ser calibrado 6 17 Permita suficiente tiempo de calentamiento para el Calibrador y el Dispositivo a ser calibrado Cuando menos 30 minutos es la recomendación normal 6 18 Ajuste una salida deseada de temperatura usando los botones arriba abajo up down de ajuste de COURSE en la carátula y la perilla de ajuste fino FINE en el panel lateral La salida de temperatura será mostrada en l...

Page 7: ...cio a Clientes al teléfono 781 890 7440 para solicitar autorización Se debe emitir un número de Autorización de Retorno RA antes de regresar cualquier producto a Extech El remitente es responsable por los cargos de envío flete seguro y empaque adecuado para prevenir daños en tránsito Esta garantía no se aplica a defectos que resulten por acciones del usuario como mal uso alambrado inapropiado oper...

Reviews: