Extech Instruments 407750 User Manual Download Page 5

Model 407750 Version 2.1 February 2004 

5

BA (Background Noise Absorber) Mode 

The Background Noise Absorber allows the user to accurately measure equipment noise 
by “eliminating” background noise. The Sound Level Meter first stores the background 
noise as a reference level. From there, when a sound is measured, the display will show 
the sound level measurement minus the background noise. To operate the meter in BA 
mode, follow these steps: 

1.  Power the meter. 

2.  Press the MAXHLD key (the MAX HOLD icon will appear on the LCD). 

3.  Press the BA key (‘F” will appear to the left of the SPL display icon). 

4.  Press the MAX HOLD key again (the MAX HOLD icon will reappear on the LCD). 

5.  The meter is now displaying the background, reference noise. 

6.  Power the device under test and note the new sound level meter reading. 

7.  If the reading changes, the new reading is the sound level of the device. If the reading 

does not change, the noise produced from the device is either equal to or less than the 
background noise. 

8.  Press the BA key again to return to the normal mode of operation. 

Auto and Manual Ranging 

The meter powers up in the Automatic Range mode. In automatic mode the meter 
automatically finds the correct range in order to produce the best accuracy. However, if it is 
desired to set the range manually, follow these steps: 

1.  Power the meter 

2.  Notice the two (2) digit number to the immediate left of the analog bargraph. This 

number is the 

low end

 of the presently selected range (see the specifications for the 

ranges). 

3.  To change the range, press the UP key to raise the range or press the DOWN key to 

lower the range. The two digit number on the left of the bargraph will change with each 
key-press.  

4.  An advantage of Manual mode is that it takes less time for the meter to take a reading. 

In Auto Range mode the meter must first locate the correct range before displaying a 
measurement. 

LCD Backlighting 

Press the BACKLIT key to illuminate the LCD. The backlight will remain on for 5 seconds 
and then automatically switch off to preserve battery life. 

Auto Power Off 

To preserve battery life, this meter has an automatic power off feature. If the unit is not 
used for approximately 20 minutes, the meter shuts off. To override this function, follow 
these steps: 

1.  From a power OFF condition, press and hold the ON/OFF and MIN/MAX keys 

simultaneously. 

2.  When ‘n’ appears on the display, release the MIN/MAX and then the ON/OFF key. 

3.  The Auto Power Off feature is now disabled. Note that the Auto Power Off feature is re-

activated the next time the meter is powered down. 

Summary of Contents for 407750

Page 1: ...User s Guide Digital Sound Level Meter Model 407750 ...

Page 2: ...d to repair or replacement of the product The warranty set forth above is inclusive and no other warranty whether written or oral is expressed or implied Introduction Congratulations on your purchase of the Extech 407750 This device measures sound level in dB and the measurement range can be set automatically or manually The 407750 offers selectable frequency weighting A and C and Time Response Fa...

Page 3: ...ct key 7 BA Background Absorber key 8 F S Fast Slow Response select key 9 DOWN 10 Backlit LCD backlighting key 11 UPPER 12 AC adaptor jack 13 Calibration screw adjust 14 AC analog output jack 15 DC analog output jack 16 RS 232 output jack Note The Battery Compartment and the Threaded Tripod mount access are located on the back of the instrument not pictured ...

Page 4: ...OW 1 second response measurements by pressing the F S key The LCD will reflect the currently selected mode Selection of Fast or Slow is determined by the application and any directives or standards related to that application For example most hearing conservation or OSHA related testing is done using SLOW and A weighting MAX HOLD The meter is capable of taking continuous measurements and only upda...

Page 5: ...ally finds the correct range in order to produce the best accuracy However if it is desired to set the range manually follow these steps 1 Power the meter 2 Notice the two 2 digit number to the immediate left of the analog bargraph This number is the low end of the presently selected range see the specifications for the ranges 3 To change the range press the UP key to raise the range or press the ...

Page 6: ...AC output is 0 707V rms full scale and the DC output is 10mV per dB The labeled 3 5mm output mini plugs are located on the bottom of the instrument RS 232 Output The meter includes an RS 232 PC interface jack This PC interface allows the meter to store and display readings on a PC as they are recorded The interface cable and software for data acquisition are sold separately Detailed instructions a...

Page 7: ...5dB 500 3 2dB 0dB 1 5dB 1 k 0dB 0dB 1 5dB 2 k 1 2dB 0 2dB 2dB 4 k 1dB 0 8dB 3dB 8 k 1 1dB 3dB 5dB Typical A Weighted Sound Levels 50HP Siren 100 130 120 110 100 90 80 70 60 50 40 30 20 10 140 0 Jet takeoff 200 dB Riveting machine Chain saw Subway 20 Boiler room Pneumatic drill 50 Freight train 100 Vacuum cleaner 10 Speech 1 Large store Large Office Small office Residence Night residential area Whi...

Page 8: ... includes a Calibration adjustment on the bottom of the meter A sound level calibrator must be used to complete a calibration Extech can supply a calibrator for this purpose Once the Calibrator is affixed on the microphone adjust the calibration screw so that the LCD sound level display matches the calibrator s output signal Extech offers repair and calibration services for the products we sell Ex...

Reviews: