Operation 

Theory of Operation 

The 407744 sound level calibrator generates a 94dB sine wave at 1 KHz. The 407766 generates a 
94dB and a 114dB sine wave at 1kHz. When a microphone from a sound level meter is inserted into 
the calibrator, the sound level meter picks up the signal and displays the signal in dB (decibel) units. 
The sound level meter  should be adju

sted (if necessary) to match the calibrator’s output signal as 

closely as possible (see diagram). 

Procedure

 

1.  Insert  the  microphone  into  the 

calibrator’s  microphone  cavity  as  shown; 

ensure  that 

it  is  fully  inserted.  Use  the  ½”  microphone  adaptor  (supplied 

attached to the calibrator) 

for ½” microphones.  

2.  On the 407744, s

lide the function switch to the ‘I’ (Power On) position. As 

soon  as  the  unit  is  powered,  the  calibration  tone  is  generated.  On  the 
407766,  slide  the  function  switch  to  the  94dB  or  the  114dB  position  as 
desired. The calibrator generates its signal as soon as the switch is set. 

3.  Read  the 

sound level meter’s display. The sound level meter should read 

94dB  or  114dB  depending  on  the 

calibrator’s  setting.  If  the  sound  level 

meter does not match the calibrator

’s signal (within specification), adjust it 

as  explained  in  the  Operation  section  (typically  a  front  or  side-panel  pot 
adjustment). Refer to diagram at right. 

4.  Slide function switch to the OFF position, and remove the microphone. 

Battery Check 

On the 407744, c

heck the condition of the battery by moving the function switch to the ‘Battery test’ 

position. The Battery status LED should illuminate if the battery is OK. If it does not light, replace the 
9V battery. On the 407766, the POWER-ON / LOW-BATTERY LED glows brightly when the battery 
is fresh and dimly when the battery is weak. 

Battery Replacement 

Lift  off  battery  compartment  cap  (shown  in  the  Description  section)  to  access  the  two  (2)  9V 
batteries. Remove the batteries and replace with new alkaline 9V batteries. Replace the cap. 
 

Never dispose of used batteries or rechargeable batteries in household waste. 
As consumers, users are legally required to take used batteries to appropriate 
collection sites, the retail store where the batteries were purchased, or wherever 
batteries are sold. 

Disposal: 

Do not dispose of this instrument in household waste. The user is obligated to 

take end-of-life devices to a designated collection point for the disposal of electrical and 
electronic equipment. 

 
 

 