Operation 

1. Power the meter by pressing the       power button. The meter will begin displaying 

sound level readings. If the LCD does not switch on, check the 9V battery located in the 
rear battery compartment. 

2. Hold the meter away from the body. 

3. View the measurement on the meter’s display. If the meter is in the autoranging mode, 

the display may briefly indicate “HI” or “LO” if the noise level is above or below the 
currently selected range. The meter will change the range as needed to display the dB 
level. 

‘A’ and ‘C’ Frequency Weighting 

Use the ‘A/C’ button to select ‘A’ or ‘C’ frequency weighting. 

With ‘A’ weighting selected, the frequency response of the meter is similar to the response 
of the  human ear. ‘A’ weighting is commonly used for environmental or hearing 
conservation programs such as OSHA regulatory testing and noise ordinance law 
enforcement. ‘C’ weighting is a much flatter response and is suitable for the sound level 
analysis of machines, engines, etc. “A” or “C” icons will appear in the display. 

Most noise measurements are performed using 'A' Weighting and SLOW Response. 

‘FAST’ and ‘SLOW’ Response Time 

Use the ‘F/S’ button to select FAST (125 ms) or SLOW (1 second) response time. Select 
FAST to capture noise peaks and noises that occur very quickly. Select the SLOW 
response to monitor a sound source that has a consistent noise level or to average quickly 
changing levels. “FAST” or “SLOW” icons will appear in the display. 

Select Slow response for most applications. 

Auto or Manual Ranging 

The meter will turn on in the auto-ranging mode and “AUTO” will be indicated on the 
display. In this mode the meter will automatically select the best range for the noise level 
being measured. If the measured dB level exceeds the range of the meter or the range of 
a selected range, “HI” will appear in the display. If the measured dB level is lower than the 
selected range, “LO” will appear in the display. 

1. Press the 

RNG

 button to manually select the range, “MANU” will be indicated on the 

display. The four ranges are: 40-70, 60-90, 80-110 and 100-130. Press the 

RNG

 button 

to step through the ranges. 

2. Press and Hold the 

RNG

 button for 2 seconds to exit the manual range mode. 

MAX HOLD 

In this mode the meter only updates the LCD when a higher reading than the one presently 
on the display is detected.  

1. Press the 

MAX HOLD

 button to enter the Max Hold mode.. The “MAX HOLD” icon will 

appear in the display.  

2. Press the 

MAX HOLD

 button again to exit this mode. 

Summary of Contents for 407730

Page 1: ...User s Guide Digital Sound Level Meter Model 407730 ...

Page 2: ...esponse Time Fast and Slow Max Hold and Max Min recording Careful use of this meter will provide years of reliable service Meter Description 1 Microphone 2 LCD Display 3 ON OFF button 4 A C weighting selection button 5 Min Max Record button 6 Range selector button 7 F S response selection button 8 Max Hold selector button 9 Calibration adjustment 10 AC analog output jack 11 Windscreen 1 Record ico...

Page 3: ...Time Use the F S button to select FAST 125 ms or SLOW 1 second response time Select FAST to capture noise peaks and noises that occur very quickly Select the SLOW response to monitor a sound source that has a consistent noise level or to average quickly changing levels FAST or SLOW icons will appear in the display Select Slow response for most applications Auto or Manual Ranging The meter will tur...

Page 4: ...button is pressed Calibration To calibrate the meter an external calibrator such as the Extech 407744 or the Extech 407766 is required in addition to a small screw driver 1 Turn the meter ON 2 Select the 80 to 110dB range 3 Select A weighting and SLOW response 4 Place the microphone into the calibrator Set the calibrator to output a 1kHz sine wave 94dB 5 Adjust the calibration potentiometer for a ...

Page 5: ...tomer Service Department at 781 890 7440 ext 210 for authorization or visit our website www extech com for contact information A Return Authorization RA number must be issued before any product is returned to Extech The sender is responsible for shipping charges freight insurance and proper packaging to prevent damage in transit This warranty does not apply to defects resulting from action of the ...

Page 6: ...t information visit our website www extech com Typical A Weighted Sound Levels 50HP Siren 100 130 120 110 100 90 80 70 60 50 40 30 20 10 140 0 Jet takeoff 200 dBA Riveting machine Chain saw Subway 20 Boiler room Pneumatic drill 50 Freight train 100 Vacuum cleaner 10 Speech 1 Large store Large Office Small office Residence Night residential area Whisper 5 0 Sound studio Threshold of hearing Copyrig...

Reviews: