407510-EU-EN-V7.2-5/12
7
Battery Replacement
The low battery indication LBT appears on the LCD when the battery runs low.
To replace the battery:
1.
Remove the meter's rubber protective cover to access the rear battery compartment.
2.
Remove the battery compartment cover using a small coin or screwdriver and remove the 9V battery.
3.
Replace the 9V battery and reinstall the cover and holster.
You, as the end user, are legally bound (
Battery ordinance
) to return all used batteries and
accumulators;
disposal in the household garbage is prohibited!
You can hand over your used batteries / accumulators at collection points in your community
or wherever batteries / accumulators are sold!
Disposal:
Follow the valid legal stipulations in respect of the disposal of the device at the end
of its lifecycle
Specifications
General Specifications
Circuit
Custom one-chip LSI microprocessor circuit
Display
Dual function 13 mm (0.5") 1999 count LCD w/ contrast adjust
Measurements Dissolved
O
2
: mg/l (milligrams per liter); Oxygen in air: %; Temperature:
°C/°F
Data hold
Data hold key freezes displayed reading
Sensor Structure
Remote Polarographic Dissolved Oxygen Sensor
Temperature sensor: Precision thermistor
Data recording
Min/Max/Avg readings stored for later recall
Auto power off
Meter turns off after 10 minutes
Sample rate
0.4 Sec (approx.)
Data Output
RS 232 PC serial interface
Operating conditions
Temperature: 0 to 50 °C (32 to 122 °F); Humidity: < 80% RH
Battery power
9V alkaline battery
Power Consumption
Approx. 6.6 mA DC
Weight
335 g / 0.74 lbs.(including batteries & probe)
Dimensions
Main instrument: 180 x 72 x 32mm (7.1 x 2.8 x1.3")
Probe: 125 x 20mm (4.9" x 0.8") length/diameter
Accessories
Carrying case, 9V battery, and 5 membranes
Optional accessories
Membranes, data acquisition software & RS232 interface cable
Electrical Specifications
Measurement Range
Resolution
Accuracy
Dissolved Oxygen
0.0 - 19.9 mg/L
0.1mg/L
+ 0.4 mg/L
Oxygen in air
0 - 100% O
2
0.1 % O
2
+ 0.7% O
2
Temperature (probe
only)
32 to 122
o
F
0 to 50
o
C
0.1
o
F/C ±1.5
o
F
+ 0.8
o
C
Temperature (automatic)
0 to 50
o
C (32 to 122
o
F)
Salt
0 to 39%
Compensation
adjustments
Altitude
0 to 3900 meters
Copyright
©
2012 Extech Instruments Corporation (a FLIR company)
All rights reserved including the right of reproduction in whole or in part in any form.