Extech Instruments 407412 User Manual Download Page 5

Model 407412 Version 2.1 July 2006 

 

airflow

Yellow
Dot

Anemometer and Temperature Operation 

1. 

Insert the Vane sensor into the meter's input jack at the top of the meter. 

2. 

Press the POWER button to turn the instrument on. 

3. 

Select Anemometer (Air Velocity) operation by placing the function slider switch to the 

vane icon    . 

4. 

Select the temperature units by pressing the °C/°F button. The LCD display will 

indicate '°C' or '°F' as selected. 

5. 

Select the air velocity unit of measure by pressing the 

UNIT button. With each press the display scrolls 
through the units (m/s, km/hr, ft/min, knots, & mile/h). 

6. 

For maximum accuracy, the yellow dot must be on the 

exhaust side of the vane as shown. 

7. 

The meter’s LCD will indicate Air Ve 

Temperature. 

Relative Humidity and Temperature Operation 

1. 

Insert the Relative Humidity probe in the input jack at the top of the meter. 

2. 

Press the POWER button to turn on the instrument. 

3. 

Select the Humidity mode of operation by placing the function slider switch to the 
RH% position. 

4. 

Select the temperature units by pressing the °C/°F button. The LCD display will 
indicate '°C' or '°F' as selected. 

5. 

Hold the probe by the handle in the area to be monitored. The meter will display 
Relative Hu Temperature. 

6. 

Note that humidity measurements take several minutes to stabilize.

 

Data Hold 

Press the DATA HOLD button to freeze the displayed reading. The DH icon will appear on 
the LCD while Data Hold is engaged. Press the DATA HOLD button again to resume 
normal operation. 

Min/Max Record Mode 

When selected, the Record mode stores the highest (Max) and lowest (Min) readings for 
later recall. To use Record mode: 

1. 

Press the RECORD button. The REC indicator will appear on the LCD display. 

2. 

Start a measurement session. 

3. 

After the measurement session, press the RECALL button to view the highest (Max) 
reading recorded since the RECORD button was pressed. The MAX indicator will 
appear to inform the user that the displayed reading is the highest value recorded. 

4. 

Press the RECALL button again to view the lowest reading encountered since the 
RECORD key was pressed (the MIN indicator will appear on the LCD to inform the 
user that the reading displayed is the lowest value recorded). 

5. 

Press RECORD again to return the meter to normal operation. All stored data will be 
cleared and the REC/MIN/MAX icons will disappear from the LCD.

 

6. 

Note that putting the meter into the Record mode defeats the Auto Power Off feature.

 

Summary of Contents for 407412

Page 1: ...User s Manual Hygro Thermo Anemometer Model 407412 ...

Page 2: ...her written or oral is expressed or implied Introduction Congratulations on your purchase of Extech s Hygro Thermo Anemometer This Heavy Duty meter measures and displays Air Velocity Temperature and Relative Humidity Temperature Air flow can be displayed in the following units of measure feet per minute meters per second miles per hour kilometers per hour and knots Temperature and RH units are dis...

Page 3: ...0 4 25 00 m s 0 01 m s 2 0 2 m sec km h kilometers hour 1 4 90 0 km h 0 1 km h 2 0 2 km hr mph miles hour 0 9 55 9 mph 0 1 mph 2 0 2 mph knots nautical miles hour 0 8 to 48 6 knots 0 1 knots 2 0 2 knots Temperature Range Resolution Accuracy 32 o F to 122 o F 0 o C to 50 o C 0 1 o F o C 1 5 o F 0 8 o C Relative Humidity Range Resolution Accuracy 10 to 70 RH 0 1 3 RH 70 to 95 RH 0 1 4 RH ...

Page 4: ... to freeze the currently displayed reading C F Press to select the temperature unit of measure RECORD Press to track the MAX and MIN readings RECALL Press to retrieve the current MAX and MIN memory FUNCTION Select Humidity left position or Air Velocity right position UNIT Select the unit of measure for Air Velocity 5 Battery compartment and tilt sand are located on the back of the meter 6 Rubber p...

Page 5: ...d 5 Hold the probe by the handle in the area to be monitored The meter will display Relative Humidity Temperature 6 Note that humidity measurements take several minutes to stabilize Data Hold Press the DATA HOLD button to freeze the displayed reading The DH icon will appear on the LCD while Data Hold is engaged Press the DATA HOLD button again to resume normal operation Min Max Record Mode When se...

Page 6: ...ined for several hours after the first appearance of the low battery indicator To replace the battery 1 Remove the meter s protective rubber holster 2 Open the battery compartment on the back of the meter using a small coin or a flat blade screwdriver 3 Replace the 9V battery close the compartment and replace the holster Relative Humidity and Temperature Calibration 1 Turn the meter on and select ...

Page 7: ... com Extech Instruments Corporation 285 Bear Hill Road Waltham MA 02451 Calibration and Repair Services Extech offers repair and calibration services for the products we sell Extech also provides NIST certification for most products Call the Customer Service Department for information on calibration services available for this product Extech recommends that annual calibrations be performed to veri...

Page 8: ...e formulae Unit of Measure Conversion Table m s ft min knots km h MPH 1 m s 1 196 87 1 944 3 6 2 24 1 ft min 0 00508 1 0 00987 0 01829 0 01138 1 knot 0 5144 101 27 1 1 8519 1 1523 1 km h 0 2778 54 69 0 54 1 0 6222 1 MPH 0 4464 87 89 0 8679 1 6071 1 Width W Height H Area A Width W x Height H CFM ft3 min Air Velocity ft min x Area ft2 CMM m3 min Air Velocity m sec x Area m2 x 60 Radius Area A x r2 W...

Reviews: