407026-EU-EN v2.6 3/14
4
Operation
Meter Power and Automatic Power Off
1. Press the POWER button to turn the meter on. If the display does not switch on, check the 9V
battery.
2. Press the POWER button again to turn the meter off.
3. The meter has an automatic power off feature that conserves battery energy. After 10 minutes
the meter automatically shuts off. To defeat this feature, press the RECORD button to put the
meter in the recording mode.
Display ZERO Calibration
Please perform a zero calibration before each use. This will ensure the highest accuracy.
1. Place the sensor cover over the light sensor, effectively blocking all light to the sensor.
2. Select the 2,000 LUX range using the RANGE switch.
3. Press the ZERO button. The reading should be zero.
4. Remove the sensor cover from the light sensor and proceed.
Selecting the unit of measure
Press the LUX/Fc button to select the desired unit of measure. The display icons will reflect the
current setting.
Selecting a light source
Select the type of lighting to be measured by pressing the LIGHT SOURCE button. The display will
indicate the lighting type icon (see icon list below).
L = Tungsten/Daylight
(use this setting when calibrating with a Tungsten lamp)
F = Fluorescent
S = Sodium
C = Mercury
Note: For Halogen and Metal Halide light use the Tungsten setting
Taking Measurements
Hold the light sensor in the area where light is to be collected. The light being measured must
encompass the entire surface of the light sensor dome (pin point lighting such as with LED light
cannot be measured).
The display will indicate the light intensity value in LUX or Foot candles (Fc). Note that since the
main display area is limited to a reading of ‘1999’, the right-most digit in the 20,000 LUX and 5,000
Fc ranges appears on the lower LCD line. In the 50,000 LUX range, the last two digits appear on the
lower LCD line
It may be necessary to adjust the display contrast due to a change in viewing angle or voltage drift.
Use the LCD Contrast adjustment located on the right side of the meter to set the preferred contrast.
Display Range Selection
Take a measurement starting at the highest range and then work down using the RANGE switch. If
the display shows dashes along the top of the reading area ( ), the input exceeds the maximum
value for the selected range; select a higher range. If the display shows dashes along the bottom of
the reading area ( ), the input is too low; select a lower range.