380820-en-GB_V1.5 3/15
8
Setting the Over Current Protection (OCP) value (Front Output)
1. Type the desired OCP (Over Current Protection) value and press the OCP button to confirm.
The OCP range: is 0.100A to 1.500A. Carefully verify this critical OCP value; this value will not
change when the unit is powered OFF.
2. When the unit is in the OPER Output mode, changing the OCP value or the Voltage value will
cause the mode to change from OPER to STBY (standby) mode. Therefore pay careful
attention to this operation.
Storing and Recalling Voltages and Frequencies (Front Output)
Often used Voltage and Frequency values can be stored in memory for quick recall. This
instrument offers 99 memory locations.
To Store:
1. Ensure that the desired voltage and frequency values are correctly set.
2. Press the STORE button (“verify STORE _” appears on display).
3. Assign a memory location number from 1 to 99 using the numeric keypad.
4. Press the V.ENTER button to confirm and save to the memory location.
5. When the unit is in the OPER Output mode, changing the OCP value or the Voltage value will
cause the mode to change from OPER to STBY (standby) mode. Therefore pay careful
attention to this operation.
To Recall:
1. Press the RECALL button (“Verify RECALL_” appears on display).
2. Select the desired memory location (1 to 99) using the numeric keypad.
3. Press the V.ENTER button to confirm. The previously stored voltage/frequency configuration
for the selected memory location will now be recalled.
Note: For an empty memory location, the LCD will display ‘BLANK’