Battery Installation 

WARNING:

 To avoid electric shock, always turn off the power and 

disconnect the test leads before opening the back to replace the fuse or 
battery. 

1.  Remove the screws securing the rear cover using a Phillips head 

screwdriver. 

2.  Lift the cover off and replace the battery observing the correct polarity. 
3.  Insert the new battery into the battery holder. 

Replace the rear cover and secure with the screws.

 WARNING: 

To avoid 

electric shock, do not operate the meter until the battery cover is in place 
and fastened securely. 

NOTE

: If your meter does not work properly, check the fuses and battery to 

make sure that they are still good and that they are properly inserted.  
 
 

Operating Instructions 

WARNING

: Risk of electrocution. High-voltage circuits, both AC and DC, are 

very dangerous and should be measured with great care. 

1.  ALWAYS turn the function switch to the OFF position when the meter is not 

in use. 

DC VOLTAGE MEASUREMENTS 

CAUTION: 

 Do not measure DC voltages if a motor on the circuit is being 

switched ON or OFF. Large voltage surges may occur that can damage 
the meter. 

1.  Set the function switch to the highest V DC position.  
2.  Insert the black test lead banana plug into the 

negative (COM) jack. 
Insert the red test lead banana plug into the positive 
(V) jack. 

3.  Touch the black test probe tip to the negative side of 

the circuit.  
Touch the red test probe tip to the positive side of the 
circuit. 

4.  Read the voltage in the display. Reset the function 

switch to successively lower V DC positions to obtain 
a higher scale reading. 

 

DC CURRENT MEASUREMENTS 

1.  Insert the black test lead banana plug into the negative 

(COM) jack. 

2.  Set the function switch to the 250mA DC position and 

insert the red test lead banana plug into the (mA) jack. 

3.  Remove power from the circuit under test, then open up 

the circuit at the point where you wish to measure 
current. 

4.  Touch the black test probe tip to the negative side of 

the circuit.  
Touch the red test probe tip to the positive side of the 
circuit. 

5.  Apply power to the circuit. 
6.  Read the current in the display. Reset the function 

switch to successively lower mA positions to obtain a higher scale reading. 

 

AC VOLTAGE MEASUREMENT 

WARNING

:  Risk of Electrocution. The probe tips may not be long enough 

to contact the live parts inside some 240V outlets for appliances because 
the contacts are recessed deep in the outlets. As a result, the reading may 
show 0 volts when the outlet actually has voltage on it. Make sure the 
probe tips are touching the metal contacts inside the outlet before 
assuming that no voltage is present. 
 

CAUTION

:  Do not measure AC voltages if a motor on the circuit is being 

switched ON or OFF. Large voltage surges may occur that can damage 
the meter. 

 

1.  Set the function switch to the highest V AC position. 
2.  Insert the black test lead banana plug into the 

negative (COM) jack. 
Insert red test lead banana plug into the positive (V) 
jack. 

3.  Touch the black test probe tip to the negative side 

of the circuit.  
Touch the red test probe tip to the positive side of 
the circuit. 

4.  Read the voltage in the display. Reset the function 

switch to successively lower V AC positions to 
obtain a higher scale reading. 

 

RESISTANCE MEASUREMENTS 

WARNING: 

To avoid electric shock, disconnect power to the unit under 

test and discharge all capacitors before taking any resistance 
measurements. Remove the batteries and unplug the line cords. 

1.  Set the function switch to the X10 or X1K position. 
2.  Insert the black test lead banana plug into the negative 

(COM) jack 
Insert the red test lead banana plug into the positive 

 

jack. 

3.  Touch the test probe tips together and adjust the OHM 

Zero Adjust knob for a “0” reading. 

4.  Touch the test probe tips across the circuit or part under 

test. It is best to disconnect one side of the part under 
test so the rest of the circuit will not interfere with the 
resistance reading. 

5.  Read the resistance on the scale 
 

dB MEASUREMENTS 

The dB scale can be used to measure the milliwatt power dissipation on a 
600

 load by measuring the voltage across a 600 

 load. An AC voltage of 

0.775Vrms across 600

 is equal to 1mW or “0”dB When converting an AC 

Voltage measurement to dB take the dB reading from the lowest arc on the 
scale plate and then add the appropriate dB correction found in the table 
below. 

 

ACV ADD 

dB 

10 0 -10 

22 

50 14  4 

36 

250 

28 

18 ~ 50 

500 

34 

24 ~ 56 

 

Maintenance 

WARNING

: To avoid electric shock, disconnect the test leads from any 

source of voltage before removing the back cover or the battery or fuse 
covers. 

 

WARNING: 

To avoid electric shock, do not operate your meter until the 

battery and fuse covers are in place and fastened securely. 

This MultiMeter is designed to provide years of dependable service, if the 
following care instructions are performed: 

1. 

KEEP THE METER DRY

. If it gets wet, dry it immediately. 

2. 

USE AND STORE THE METER IN NORMAL TEMPERATURES.

 

Temperature extremes can shorten the life of the electronic parts and 
distort or melt plastic parts. 

3. 

HANDLE THE METER GENTLY AND CAREFULLY.

 Dropping it can 

damage the electronic parts or the case. 

4. 

KEEP THE METER CLEAN.

 Wipe the case occasionally with a damp 

cloth. DO NOT use chemicals, cleaning solvents, or detergents. 

5. 

USE ONLY FRESH BATTERIES OF THE RECOMMENDED SIZE AND 
TYPE. 

Remove old or weak batteries so they do not leak and damage the 

unit. 

6. 

IF THE METER IS TO BE STORED FOR A LONG PERIOD OF TIME

the batteries should be removed to prevent damage to the unit. 

 

UL LISTED 

The UL mark does not indicate that this product has been evaluated for the 
accuracy of its readings. 
 

Warranty 

EXTECH INSTRUMENTS CORPORATION warrants this instrument 
to be free of defects in parts and workmanship for one year from date 
of shipment (a six month limited warranty applies on sensors and 
cables). If it should become necessary to return the instrument for 
service during or beyond the warranty period, contact the Customer 
Service Department at (781) 890-7440 ext. 210 for authorization. 

Return Authorization (RA) number must be issued before any 
product is returned to Extech.

 The sender is responsible for 

shipping charges, freight, insurance and proper packaging to prevent 
damage in transit. This warranty does not apply to defects resulting 
from action of the user such as misuse, improper wiring, operation 
outside of specification, improper maintenance or repair, or 
unauthorized modification. Extech specifically disclaims any implied 
warranties or merchantability or fitness for a specific purpose and will 
not be liable for any direct, indirect, incidental or consequential 
damages. Extech's total liability is limited to repair or replacement of 
the product. The warranty set forth above is inclusive and no other 
warranty, whether written or oral, is expressed or implied.

 

Repair and Calibration Services 

Extech offers complete repair and calibration services

 for all Extech 

products. For periodic calibration, NIST certification or repair of any Extech 
product, call customer service for details on services available. Extech 
recommends that calibration be performed on an annual basis to insure 
calibration integrity. 

 

Copyright © 2006 Extech Instruments Corporation. All rights reserved 

including the right of reproduction in whole or in part in any form. 

38070   V3.0     01/07 