Stroke Measurement
Press [MODE] until you enter the Stroke Measurement mode. The display will be as follows.
In the Stroke Measurement mode, the duration of 3 strokes is taken and the stroke frequency (counts per
minute) is calculated. As the stroke frequency falls out of the range of 10 - 180, "Err" will appear to show the
error condition.
Press [START] to start the measurement. During the 1st second, 180.0 will be shown and flashing as below.
At the end of the first second, the display will become steady.
If measurement is stopped in the first second (stroke frequency> 180), error signal " Err "will be shown. After
the first second has elapsed, the instantaneous measurement reading will be shown. When you press [STOP],
the display will show the steady stroke measurement reading. If the measurement lasts for more than 18
seconds (stroke frequency < 10), an error signal "Err" will be displayed. When the measurement is stopped,
press [RESET] to reset the displayed reading to "0.0