background image

 

 
User's Guide 

 
 
 
 
 

Oyster pH - Conductivity - TDS - Temperature Meters 
 
Models  
34135-A 
341350 
341350-P 
34145A 
341450 
341450-P 

 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 

INTRODUCTION 

 
Congratulations on your purchase of the Extech's Oyster Series meter or meter kit. These 
devices measure pH/Conductivity plus TDS or Temperature. These meters are intended for 
routine laboratory and field testing. Accurate measurements are provided in a battery 
operated, portable meter with a hinged display that can be adjusted to any viewing angle. 
 

Model 34135-A

 measures pH, Conductivity and TDS. The ‘A’ indicates ‘meter-only’.  

Models 341350 and 341350-P

 are kits which include the 34135A and all the necessary 

accessories. The ‘P’ indicates Polymer type electrode (341350 has glass electrode). 
 

Model 34145-A

 measures pH, Conductivity & Temperature. ‘A’ indicates ‘meter-only’. 

Models 341450 and 341450-P

 are kits which include the 34145A and all the necessary 

accessories. The ‘P’ indicates Polymer type electrode (341450 has glass electrode). 

 

Test Equipment Depot - 800.517.8431 - 99 Washington Street Melrose, MA 02176

FAX 781.665.0780 - TestEquipmentDepot.com 

 

99 Washington Street 
Melrose, MA 02176 

Phone

 781-665-

1400

Toll Free 1-800-517-8431

 

Visit us at www.T

estEquipmentDepot.com

Back to the 

Extech 341350

 Product Page

Summary of Contents for 34135-A

Page 1: ...s pH Conductivity and TDS The A indicates meter only Models 341350 and 341350 P are kits which include the 34135A and all the necessary accessories The P indicates Polymer type electrode 341350 has glass electrode Model 34145 A measures pH Conductivity Temperature A indicates meter only Models 341450 and 341450 P are kits which include the 34145A and all the necessary accessories The P indicates P...

Page 2: ...anges Low Battery Indication Automatic Dimensions Weight 3 75 x 4 25 x 2 96 x 108 x 45mm 12 oz 340g Accessories supplied 9V battery plus 341350 kit pH electrode pH buffers bottle case glass conductivity cell 341350 P pH electrode pH buffers bottle case polymer conductivity cell 341450 kit pH electrode pH buffers bottle case glass conductivity cell stainless steel temperature probe 341450 P pH elec...

Page 3: ...l measurements have been taken unplug the electrode or cell and rinse in distilled water shake and store as described below Conductivity TDS Calibration The conductivity cell calibration should be performed frequently to ensure accuracy 1 Connect the conductivity cell to the input socket on the side of the meter 2 Prepare a conductivity solution of approximately the same value as the solution to b...

Page 4: ...ts bottle Use a pH 4 buffer solution or tap water as the storage medium 3 If bubbles are seen in the bulb area hold the electrode by its cap and shake downwards until bubbles are removed 4 To improve speed of response vigorously stir the electrode in the sample buffer or rinse solution 5 After exposure to a sample buffer or rinse solution shake the electrode with a snap motion to remove residual d...

Page 5: ...ratic LCD display Samples have low ionic strength lacks salt e g distilled de ionized boiled lake water high pressure For each 50 ml of sample add 1 drop 50uL of SAT KCL No alteration in pH will occur by inert KCL MAINTENANCE Conductivity Cell Simulator The meter is supplied with a cell simulator which can be used in place of the conductivity cell to provide a means to verify the operation of the ...