Using the Optional Power Meters and VFL
Power Meter and VFL
11
Measuring Power or Insertion Loss
To perform power or insertion loss measurements manually:
1.
If necessary, perform an offset nulling (see
2.
Check your fibers and clean them properly.
3.
For insertion loss measurements, reference your power meter to a light
source (see
Setting Reference Values on Your Power Meter
then deactivate the light source.
4.
If you have used a single reference patchcord, disconnect it
from the
power meter port only
, then attach a second reference patchcord to
the power meter.
OR
If you have used two reference patchcords, disconnect both of them at
the bulkhead.