SCPI Command Reference
Tunable Laser Source
95
T200S Specific Commands
OUTPut:SYNChronization[:STATe]
Description
This command controls the activation of the output signal
generation on the
Trig out
BNC connector located on the rear
panel.
At *RST, this value is set to OFF.
Syntax
:OUTPut:SYNChronization[:STATe]<wsp><state>
Parameter(s)
state
:
State of the Trig out output. The allowed values are:
³
OFF|0 (default): the output is closed, which means that no
signal is output from the BNC port.
³
ON|1: the output is open.
Example(s)
OUTP:SYNC:STAT OFF (closed)
See Also
OUTPut:SYNChronization[:STATe]?
OUTPut:SYNChronization[:STATe]?
Description
This query indicates if the output signal generation is active for the
Trig out
BNC connector located on the rear panel.
At *RST, this value is set to OFF.
Syntax
:OUTPut:SYNChronization[:STATe]?
Parameter(s)
None.
Response Syntax
<state>
Response(s)
state:
State of the Trig out output:
³
0 (default): the output is closed, which means that no signal is
output from the BNC port.
³
1: the output is open.
Example(s)
OUTP:SYNC ON
OUTP:SYNC? returns 1 (open)
See Also
Summary of Contents for T200S
Page 1: ...User Guide T200S Tunable Laser Source www EXFO com...
Page 18: ......
Page 48: ......
Page 66: ......
Page 144: ......