IQS-12001B
Cable Assembly and Component Test System
SPECIFICATIONS
a
Singlemode
b
(1310/1550 nm)
Singlemode
b
(1490/1625 nm)
Multimode
c
Testing time
d
(s)
< 6
< 6
< 6
Cable assembly length (m)
1.8 to 1500
1.8 to 1500
1.8 to 500
Insertion loss measurement uncertainty
± 0.03
e
± 0.03
e
± 0.07
f
Insertion loss measurement stability
g
(dB)
± 0.004
± 0.004
± 0.015
Reflectance range (dB)
−30 to −75
−30 to −75
−10 to −50
Reflectance measurement uncertainty (dB)
±1 (−30 to −70)
±1 (−30 to −70)
± 1.2 (−10 to –30)
± 1.5 (−30 to –40)
±2.2/±1.7 (−70 to −75)
±1.7 (−70 to −75)
± 1.6 (−40 to –43)
± 2.9 (−43 to –50)
Reflectance measurement repeatability
h
(dB)
±0.1 (−30 to −65)
±0.1 (−30 to −65)
± 0.2 (−10 to –30)
±0.2 (−65 to −70)
±0.2 (−65 to −70)
± 0.4 (−30 to –40)
±0.4 (−70 to −75)
±0.5 (−70 to −75)
± 0.6 (−40 to –43)
± 1.8 (−43 to –50)
Wavelengths (nm)
1310/1550
1490/1625
850/1300
Output channels
Up to 32 channels
Up to 32 channels
Up to 32 channels
Test method
End-to-end/bidirectional
End-to-end/bidirectional
End-to-end/bidirectional
Notes
a. Calculated and measured with the following considerations:
• Based on recommended procedure for ORL measurements
• The FOA-300 series’ uncertainty is included in all uncertainty values
• Although tests are possible with other fiber-optic adapters, these specifications are only valid with the FOA-300 series
• Cable assemblies <1.8 m can be tested using a non-reflective termination
• With 30 minute warm-up time
b. For SMF-28 fiber. At 23 ºC ± 1 ºC in High-Sensitivity Reflectance mode with no switch.
c. At 850 nm and 1300 nm. For 62.5/125 µm fibers. At 23 °C ± 1 °C.
Uncertainty due to launching conditions not included.
d. Calculation and storage time for simplex, dual-wavelength IL/reflectance end-to-end measurements (Standard Reflectance Sensitivity mode, 500 meter range);
does not include referencing and connection time.
e. For simplex measurements using FOA-322 adapter. Uncertainty for MTP to MTP, MT-RJ to MT-RJ, MTP to fanout and MT-RJ to fanout is ± 0.06 dB
(reported with a level of confidence of 95 %). This does not include uncertainties due to connector, connector adapter or switch PDL.
f. For simplex measurement using FOA-322 adapter. Uncertainty for MTP to MTP, MT-RJ to MT-RJ, MTP to fanout and MT-RJ to fanout is ± 0.08 dB
(calculated for a level of confidence of 95 %). Does not include uncertainties due to connector, connector adapter or uncertainties due to mode dependence of the switch.
g. For a stable connection, over 15 minutes, at constant temperature. The stability is expressed as ± half the difference between the maximum and minimum values measured
during the period.
h. For a stable connection, over 10 measurements.
GENERAL SPECIFICATIONS
Communication interfaces
General-purpose interface bus (GPIB)
Ethernet (10/100 Base-T)
RS-232
Temperature
Operating
0 °C to 40 °C
(32 °F to 104 °F)
Storage
−40 °C to 60 °C
(−40 °F to 140 °F)
Relative humidity
80 % maximum, non-condensing
Laser Safety
21 CFR 1040.10
CLASS 1 LASER PRODUCT
IEC 60825-1:2007
CLASS 1 LASER PRODUCT
Standard Accessories
Expansion units as required, user guide, certificate of compliance, custom report, label generator, IQS-12001B software and EUI-89 FC narrow key connector.