
Acquisition Data
120
FTB-5600
The FTB-5600 takes care of this issue: it automatically computes all of the
sampling steps required between the span provided. Only the value at the
end of the link (the PMD scale), and the minimum value desired (the
residual PMD) need to be set.
Note:
The PMD sampling steps overlap by 3.
In the example below, the chosen scale is 8 ps and the residual PMD of
60 fs. We can see each scales with their saturation level and how they are
used to compute the overall PMD curve.
Scale (ps)
Step (GHz)
Residual PMD (ps) Saturation (ps)
8
12.5
1.6
19.4
2.67
37.5
0.533
6.5
0.889
112
0.178
2.15
0.296
337
0.06
0.7
Saturation at 6.5 ps
Saturation at 2.15 ps
Saturation at 0.7 ps
Summary of Contents for FTB-5600
Page 1: ...User Guide Distributed PMD Analyzer for FTB 500 FTB 5600...
Page 12: ......
Page 14: ......
Page 77: ...Managing Results Distributed PMD Analyzer 71 Editing Sections Trace after treatment...
Page 106: ......
Page 110: ......
Page 116: ......
Page 118: ......