Controlling the Source Remotely
26
Setting the Flow Control
Setting the Flow Control
The flow control parameter applies only to RS-232 communication. This
parameter determines the type of serial communication used.
Choose
Soft
if you want the speed of data transmission to match the speed
at which the device can process it. This enables the computer and the
CD/PMD Analyzer Source to stop each other from transmitting by sending a
control character (Xoff). They will also be able to restart the transmission
by sending another control character (Xon). This is known as a software
handshake, a procedure requiring a special cable.
To set a flow control:
1.
Press the Setup key.
2.
Use the up/down or left/right arrow keys to select
Flow Ctrl
. The
current setting is displayed.
3.
Press ENTER, then use the up/down arrow keys to select the type of
flow you want. None means no flow control. Soft
allows the unit or
computer controlling it, to turn the data transmission on or off.
4.
Press ENTER to confirm.
Refresh Rate
Backlight
Contrast
Video Mode
Exit
8 Hz
ON
STD
RS232 / GPIB
GPIB Addr.
Baud Rate
Flow Ctrl
RS232
N.A.
19200
Soft