Controlling the Source Remotely
CD/PMD Analyzer Source
27
Communication Parameters
Communication Parameters
The communication parameters are used to set the communication port.
Note:
EOS means “End of String.” EOI means “End or Identify.”
For GPIB Communication
Terminate Read on EOS
Yes
Set EOI with EOS on Writes
Yes
Type of compare on EOS
8 bits
EOS byte
0Ah
Send EOI at end of Writes
Yes
GPIB primary address
See
Setting the GPIB Address
on
page 24
GPIB secondary address
None
For RS-232 Communication
EOS bytes
0Ah
Baud rate
1200/2400/4800/9600/ 19200 bps
Parity
None
Data bits
8 bits
Stop bits
1 bit
Flow control
Software (Xon/Xoff) or None
Activation
See
Setting the Remote Command
Mode
on page 23