Troubleshooting
60
FLS-2300B
FLS-2300B Error Messages
-34
Command overflow:
Too many commands were sent
to the unit at a time.
Wait until the unit is done before
sending more commands.
-40
FIFO not ready for reading:
The unit’s FIFO is not ready for
reading. Commands sent will be
ignored.
If problem persists, call EXFO for
assistance.
-60
Laser over-current:
A problem occured with the
current going to the laser.
Call EXFO for assistance.
-64
EEPROM error:
The EEPROM was not detected.
Call EXFO for assistance.
-65
EEPROM checksum error:
A Checksum error was detected
by the unit.
Call EXFO for assistance.
-68
Temperature error:
The operating temperature of
the unit is from 0 °C to 40 °C
(32 °F to 104 °F).
Make sure the ambient
temperature is comprised in the
specified temperature range.
-69
FPGA problem:
A FPGA problem was detected.
Call EXFO for assistance.
-73
Supply voltage error:
Internal voltage values are out of
the operation range.
Call EXFO for assistance.
-74
Laser pump failure:
A problem occured in the output
(optical) power of the laser
pump.
Call EXFO for assistance.
Warning/Error
Number
Description
Recommended Action
Troubleshooting
60
FLS-2300B
FLS-2300B Error Messages
-34
Command overflow:
Too many commands were sent
to the unit at a time.
Wait until the unit is done before
sending more commands.
-40
FIFO not ready for reading:
The unit’s FIFO is not ready for
reading. Commands sent will be
ignored.
If problem persists, call EXFO for
assistance.
-60
Laser over-current:
A problem occured with the
current going to the laser.
Call EXFO for assistance.
-64
EEPROM error:
The EEPROM was not detected.
Call EXFO for assistance.
-65
EEPROM checksum error:
A Checksum error was detected
by the unit.
Call EXFO for assistance.
-68
Temperature error:
The operating temperature of
the unit is from 0 °C to 40 °C
(32 °F to 104 °F).
Make sure the ambient
temperature is comprised in the
specified temperature range.
-69
FPGA problem:
A FPGA problem was detected.
Call EXFO for assistance.
-73
Supply voltage error:
Internal voltage values are out of
the operation range.
Call EXFO for assistance.
-74
Laser pump failure:
A problem occured in the output
(optical) power of the laser
pump.
Call EXFO for assistance.
Warning/Error
Number
Description
Recommended Action
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
Summary of Contents for FLS-2300B
Page 9: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...
Page 13: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...
Page 21: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...
Page 33: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...
Page 53: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...
Page 67: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...
Page 95: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...