![Exergen LXTA-5000 Reference Manual Download Page 27](http://html.mh-extra.com/html/exergen/lxta-5000/lxta-5000_reference-manual_2449821027.webp)
27
Care and Maintenance
•
Accuracy Limits: Comparison between the reference instrument(s) and the instrument being verified
should be within ±0.4ºF (0.2ºC) for acceptable limits. If not, repeat the process. In the event they
still differ by more than the acceptable limits, call Exergen for repair or replacement of the failed
instrument.
V
VV
VVerif
erif
erif
erif
erifier Specif
ier Specif
ier Specif
ier Specif
ier Specifications:
ications:
ications:
ications:
ications:
Power Source
Power Source
Power Source
Power Source
9-volt battery, or 9-volt power supply
Battery Life
Battery Life
Battery Life
Battery Life
Approximately 1 hour continuous use
Low Voltage Indicator
Low Voltage Indicator
Low Voltage Indicator
Low Voltage Indicator
Red LED shuts off when battery voltage drops below ~5
volts
Temperature Range
Temperature Range
Temperature Range
Temperature Range
97-104 °F (36-40 °C)
Cleaning
Cleaning
Cleaning
Cleaning
Wipe down with alcohol or any hospital approved
disinfectant. Do not immerse.
R
R
R
R
Repair
epair
epair
epair
epair
If repair is required:
•
Contact Exergen at (617) 923-9900 for a Return Materials Authoriza-
tion (RMA) Number.
•
Mark the RMA number on the outside of your package and packing
slips.
•
Include a description of the fault if possible.
•
Send the instrument freight/postage prepaid to:
Exergen Corporation
51 Water Street
Watertown, MA 02472
•
The instrument will be returned freight/postage prepaid.