12
EXCELL PRECISION CO., LTD.
ZSME300000276
02005324
2-1-3 Hi/Lo/OK Function Setting
Select
in the General Function setting mode
to set the
Hi/Lo/OK function. This function is available in all unit modes. In one specific unit mode,
enter
to set the Hi/Lo/OK values.
When the high limit and low limit are both set as “0”, the Hi/Lo/OK function is disabled.
ACTIONS
DISPLAY
NOTE
LED displays the last status
Press
TARE
key
Key in the desired high
limit value
Press
TARE
key
Key in the desired low limit
value
Press
TARE
key
Key in the parameter
setting for alarm
Press
TARE
key
LED displays the last status
LED displays the last status
.
.
.
.
Alarm setting
(a) (b) (c)
The definition of a, b, c positions:
(a)
⇒
1 = beep sound on
0 = beep sound off
(b)
⇒
1 = beep sound on when stable
0 = beep sound on when unstable
(c)
⇒
1 = beep sound on when the weight is
between high and low limits
0 = beep sound on when the weight is not
between high and low limits and
higher than 10d
ON/ZERO
key
⇒
Upward key (from 0 to 9)
UNITS
key
⇒
Downward key (from 9 to 0)
TARE
key
⇒
Move the cursor to right
NET/GROSS
key
⇒
Move the cursor to left.
Key in the parameter and then press
TARE
key to go back to the
Advanced function setting menu