X-Cite® XYLIS User Guide
035-00683R Rev1
Figure 9 Location and Removal of Fuse Tray in AC Receptacle
b.
If either or both of the fuses are open, replace with the same type (5A, 250V).
c.
Note: To determine if a fuse is intact (i.e. OK) or open (i.e. blown), remove the fuse
from the tray and check with a multi-meter set to resistance (Ω). An intact fuse will read
“0Ω” (or another very low value), an open fuse will have an extremely high Ω reading.
d.
Note: Fuses that chronically need replacing are usually a signal that something is wrong
and Technical Support should be contacted.
3.
SpeedDIAL Check:
a.
Verify that the pins on speedDIAL connector are straight.
b.
Verify that speedDIAL is plugged securely into main unit.