22
X-Cite® mini+ User Guide
035-00587R rev. 5
Interval
Description
Time (µs)
T1
Delay time, TTL trigger on to LED on
100 @ 100% power
T2
Delay time, TTL trigger off to LED off
30
5.2.4
TTL Signal and LED Status
TTL
LED Status
High
ON
Low
OFF
5.2.5
TTL Input Specifications
a.
Connector type: BNC (female port)
b.
Maximum low level : +0.8V
c.
Minimum high level : +2.2V
d.
Maximum high level: + 5.5V
e.
Typical input current: 800µA
5.3
Foot Pedal Control (Optional)
For hands-free operation, a foot pedal can be used to manually turn the LED on and off. The foot pedal
is an optional accessory and can be purchased separately from Excelitas Technologies.
1.
Installing the foot pedal
a.
Locate the “FOOT PEDAL IN” / “FP IN” input port on the rear panel of X-Cite® mini+.
b.
Insert foot pedal plug.
c.
Place foot pedal on floor.
2.
Operating the foot pedal
a.
Press and release pedal with your foot to toggle LED on and off.
b.
On/off status of LED will be shown on speedDIAL display.
c.
If desired, the foot pedal can be used in combination with speedDIAL (e.g. turn on with
speedDIAL, turn off with foot pedal.) The ON/OFF status of the system will be reflected
on speedDIAL’s home screen whether it is changed via the foot pedal or by any other
means.