![Evoqua WALLACE & TIERNAN DEOX/2000 Instruction Manual Download Page 12](http://html1.mh-extra.com/html/evoqua/wallace-and-tiernan-deox-2000/wallace-and-tiernan-deox-2000_instruction-manual_2447568012.webp)
W3T331759 / WT.050.585.003.UE.IM.1014
8
MICRO/2000® AND DEOX/2000® MODULE
full scale (whichever is greater)
DEOX/2000® - 0.01 mg/l or 5 %
full scale (whichever is greater)
Protection Category:
CE Version Enclosure - (IP 66), CE
(NEMA 4X)
Power Supply:
230V ± 10%, 50/60 Hz, 15W;
Fuse - 200 mA, 5 x 20 mm, fast acting
Operating Conditions:
Ambient temperature 2 to 52°C
(35 to 125°F )
Sample Temperature -3 to 52°C
(26 to 125°F )
Storage Temperature -16 to 70°C
(4 to 158°F )
1.1.1 MICRO/2000® and DEOX/2000® flow block assembly
Housing
Facade Dimensions:
(W x H x D) mm: 330 x 330 x 254
Weight: approx.:
9.1 kg
Connections
Sample water:
6mm OD (¼") hose
Thread connection:
13mm (½")
Flow control valve (to sample inlet water line accessories)
Flow rate:
MICRO/2000® - approx. 3.8 to 11.4 l/m
(1 to 3 US gpm)
DEOX/2000® - approx. 7.6 to 19 l/m
(2 to 5 gpm)
Control range:
0.2 to 4.0 bar (3 to 58 psi)
Back-pressure:
0 bar (0 psi) (free drain)
Sample water
Water quality:
ultra pure, potable, industrial, municipal
and industrial waste, dechlorinated waste
(DEOX/2000®)
Sample water
temperature:
max. 50 °C (122 °F)
Summary of Contents for WALLACE & TIERNAN DEOX/2000
Page 4: ......
Page 10: ...W3T331759 WT 050 585 003 UE IM 1014 6 MICRO 2000 AND DEOX 2000 MODULE...
Page 30: ...W3T331759 WT 050 585 003 UE IM 1014 26 MICRO 2000 AND DEOX 2000 MODULE...
Page 46: ...W3T331759 WT 050 585 003 UE IM 1014 42 MICRO 2000 AND DEOX 2000 MODULE...
Page 74: ...W3T331759 WT 050 585 003 UE IM 1014 70 MICRO 2000 AND DEOX 2000 MODULE...