![Evikon PluraSens E2660-CO-NO User Manual Download Page 18](http://html1.mh-extra.com/html/evikon/plurasens-e2660-co-no/plurasens-e2660-co-no_user-manual_2446125018.webp)
E2660-CO-NO
Rev 29.06.2021
0x0006
6 / 40007 RW Response delay, ms
10...255
10
0x0007
7 / 40008 RW Stop bits, parity bit *
1 – No parity bit, 1 stop bit
(default after factory reset)
2 – No parity bit, 2 stop bits
3 – Odd parity, 1 stop bit
4 – Even parity, 1 stop bit
NOTE: 3 and 4 are available
starting from the Software
version 0x204 (dec. 516)
1
0x0008
8 /40009
R
Last error code
1...255
-
0x0011
17 / 40018 RW
Technological:
age of last data in seconds
(read) / restart (write)
0...65535 s (read),
writing 42330 restarts the
instrument
-
0x00A5 165 / 40166 RW
Zero adjustments for gas
data, ADC
-32000...+32000 ADC units
0
0x00A6 166 / 40167 RW Slope adjustment for gas
data
1...65535
512
0x00A7 167 / 40168 RW The change rate limit for
gas data, ppm gas unit/s
1...32000, 0 - no limit
0
0x00A8 168 / 40169 RW Integrating filter time
constant, s
1...32000 (seconds),
0 - no filter
0
Addr
Reg / MHR RW Description
Supported values (dec)
Default
0x00C9 201 / 40202 RW
Parameter tied to analog
output 1
0 – None
2 – 1st Gas concentration
3 – 2nd Gas concentration
9 – Forced Modbus control,
value set in MHR / 40204
2
0x00CA 202 / 40203 RW
Parameter tied to analog
output 2
0 – None
2 – 1st Gas concentration
3 – 2nd Gas concentration
9 – Forced Modbus control,
value set in MHR / 40205
3
0x00CB 203 / 40204 RW
Forced value for analog
output 1***
0...1000
(0.0%...100.0% of output
scale)
0
0x00CC 204 / 40205 RW
Forced value for analog
output 2***
0...1000
(0.0%...100.0% of output
scale)
0
0x00D3 211 / 40212 RW Parameter tied to relay
0 –none
2
18