Evikon PluraSens E2638-VOC User Manual Download Page 1

Volatile Organic Compounds
Highly volatile organic liquids. The definitions of VOC may vary depending on
application field and country. E2638-VOC is intended to detect vapours of solvents
(acetone, benzene, ethanol, ethyl acetate, toluene, xylenes etc.
See Annex 2 for more information on the properties of various gases.

Installation guidelines
(See Installation and connections section for general information.)
Solvent  vapours  are  heavier  than  air  and  tend  to  sink.  Consider,  if  the  vapours  are
heated or not, if the detector is used for fire safety (LEL range) or air quality control
(0...100 to 0...1000 ppm range) etc.
Calibration

E2638-VOC  detectors-transmitters  have  been  calibrated  by  Manufacturer  with
standard  gas  mixtures  before  delivery.  Provided  that  the  sensor  is  used  under
moderate  conditions,  field  recalibration  is  recommended  once  a  year.  Please
contact your dealer for more information.

E2638-VOC_UM_EN 

         Rev 25.05.2020

Specifications

Sampling method

diffusion

Sensor type

metal oxide semiconductor

Default calibration

toluene

Typical detection range

0...100% LEL

up to 0...1000 ppm

Resolution / digital unit

0.1% LEL

1 ppm

Response time T90

< 120 s

Sensor lifetime

> 5 years

Calibration interval

12 months

Signal update

every 1 second

Power supply options

11...30 VDC, 24 VAC

or 90...265 VAC (with mains power unit)

Power consumption

< 2 VA

Analog outputs

2 × 4-20 mA / 0-10 V, user settable

Outputs assignment

OUT1 2 gas; OUT2 2 gas

NOTE for LCD version only output 1 is available

Load resistance

R

L

 < (Us - 2 V) / 22 mA for 4-20 mA

R

L  

> 250 kOhm for 0-10 V mode

Digital interface

RS485, Modbus RTU protocol; no galvanic isolation

Enclosure

grey ABS, wall mount, protection class IP65

Dimensions

H90 × W145 × D55 mm

(overall dimensions H140 × W145 × D55 mm)

CE marking

according to 2014/30/EU and EN61326-1 requirements

Operating conditions

-30...+70 °C; <95% RH non-condensing, 0,9...1,1 atm 

explosion safe areas; normal ambient oxygen level

avoid exposure to corrosive gases or silicone

containing products

Relay option

Output relays

2 × SPST relays (closing contact),

250 VAC / 30 VDC, 5 A max

Alarm setpoints
(set/release)

RE1 (LOW): 20/16 % LEL;

RE2 (HIGH): 50/40% LEL

RE1 (LOW): 100/80 ppm;

RE2 (HIGH): 300/240 ppm

(for 0...500 ppm range)

Signalling options
Visual

red and green LEDs

Acoustic alarm

buzzer 85 dB

LCD indicator option
Operating temperature

0...+50 °C

Display dimensions

72 × 36 mm

Number of digits

3.5 7-segment

Character height

14 mm

Other features

backlight

Other options

Remote probe

protection IP65, default cable length 3.0 m;

max height 80 mm, max diameter 65 mm

         Evikon MCI OÜ

                Teaduspargi 7/9, Tartu

[email protected]

         Tel. +372 733 6310 

  50411 Estonia

www.evikon.eu

PluraSens

®

Solvent Vapours Detector-Transmitter

E2638-VOC

User Manual