background image

E2630-HFC

Rev 23.03.2021

Product description

E2630 series gas detectors are compact and easy-to-use instruments. The devices utilize
novel fully calibrated and temperature compensated gas sensors with excellent
repeatability, stability, and long lifetime.

Two relays RE1 and RE2 with switching contacts can be used to control alarm sirens,
ventilation fans, shut-off valves, or other actuators. The devices are equipped with a visual
and acoustic alarm.

The version of your detector is marked on the package.

Safety requirements

Misuse will impair the protection of the product. Always adhere to the safety provisions
applicable in the country of use.

Do not perform any maintenance operation with the power on. Do not let water or foreign
objects inside the device.

Removal of the PCB from the enclosure voids the warranty. Do not touch the electronic
components directly, as they are sensitive to static electricity.

Connection diagrams can be found in the connections section. The device might not
perform correctly or be damaged if the wrong power supply is connected.

External circuits connected to the equipment should have sufficient insulation rating
according to the environmental conditions and equipment power.

A disconnecting device that is marked as such and easily accessible should be included in
the installation of this product.

Operating conditions

The device should be used both in a non-hazardous indoor area and in a basic
electromagnetic environment, where the latter is defined in EN 61326-1. Avoid strong
mechanical shock and vibrations. Avoid corrosive atmosphere and areas highly
contaminated with dust, oil mist, etc. Keep the instrument away from direct sunlight. A
sudden temperature or humidity change might affect the sensitivity of the sensor.

When stored without powering in normal air for a long period, or in an environment
contaminated with organic vapors or volatile oils, the sensor may show a reversible drift in
resistance according to the environment.

6

Summary of Contents for E2630-HFC

Page 1: ...Refrigerant Detector E2630 HFC User Manual...

Page 2: ...duct description 6 Safety requirements 6 Operating conditions 6 Installation guidelines 7 Mounting dimensions 8 Connections 9 Operation 10 Sensor probe handling 11 Maintenance 12 Calibration 12 Delive...

Page 3: ...e environment Now hydrofluorocarbons are the most commonly used halocarbon refrigerants Hydrofluoroolefins HFO is the last generation of refrigerants more environmentally friendly than HFCs but modera...

Page 4: ...21 asphyxiation in confined spaces At higher temperatures 250 C decomposition products may include hydrofluoric acid HF and carbonyl halides An escape of refrigerant through a leak may damage the refr...

Page 5: ...face UART Relay outputs 2 SPDT max 5 A 30 VDC 250 VAC Alarm signaling Buzzer 2 kHz 85 dB red LED LEDs Green red operation fault red gas alarm Enclosure Grey ABS wall mount protection class IP65 Dimens...

Page 6: ...o static electricity Connection diagrams can be found in the connections section The device might not perform correctly or be damaged if the wrong power supply is connected External circuits connected...

Page 7: ...a of the device however the coverage area for a detector is usually between 2 5 to 5 meters radius For early leakage detection install the sensor as close as possible to the potential leakage sources...

Page 8: ...E2630 HFC Rev 23 03 2021 Mounting dimensions 8...

Page 9: ...5 mm2 We recommend to strip the wire end by 5 6 mm and use the wire end sleeves To connect the wire loosen the screw insert the wire end into the terminal hole and tighten the screw The outer diameter...

Page 10: ...gas concentration exceeds the LOW alarm setpoint the bottom red LED starts flashing at a rate of 1 Hz and the relay RE1 switches over The first alarm stops automatically if the gas concentration drop...

Page 11: ...lter Place a new filter into the cap and tighten it again NOTE Never stab or press the filter near its center where the sensor is located since this may damage the sensor Do not remove the filter as i...

Page 12: ...e Manufacturer with standard gas mixtures before delivery Provided that the sensor is used under moderate conditions field recalibration is recommended every 12 months Please contact your dealer for m...

Page 13: ...Manufacturer will at its option either repair or replace a product that proves to be defective This warranty is void if the product has been operated in conditions outside ranges specified by the Manu...

Reviews: