Evikon E2618-H2S User Manual Download Page 2

Return to default settings

To reset the device’s Slave ID, baudrate and sbit number to factory settings,
proceed as follows:
1. De-energize the device
2. Connect the J3 jumper
3. Turn on the device
4. De-energize the device
5. Disconnect the J3 jumper
6. Turn on the device

Calibration

E2618-H2S has been calibrated by Manufacturer with standard gas mixture before
delivery. The electrochemical gas sensor exhibits high stability and ~2 year
lifetime. However, as the gas sensor is directly exposed to environment, the
instrument requires a field recalibration with a portable calibration kit at least every
6 month. Please contact your distributor for more information.

Delivery set

– Hydrogen suldfide transmitter E2618-H2S or E2618-H2S-DM
– Mounting accessories:

4 screws with plastic dowels for wall mount version
25mm cable gland for duct mount version.
fixing clamp for remote probe versions

Warranty

This product is warranted to be free from defects in material and workmanship for a
period of one year from the date of original sale. During this warranty period
Manufacturer will, at its option, either repair or replace product that proves to be
defective. This warranty is void if the product has been operated in conditions
outside ranges specified by Manufacturer or damaged by customer error or
negligence or if there has been an unauthorized modification.

E2618-H2S_UM_EN

        Rev 04.05.2016

General technical data

Sensor type

long-life electrochemical gas sensor

Detected gase

Hydrogen sulfide H

2

S

Calibration

Hydrogen sulfide H

2

S

Sampling method

diffusion

Detection range options

standard 0...100 ppm (500 ppm peak) H

2

S

high 0...1000 ppm (10 000 ppm peak) H

2

S

Resolution

0,1 / 1 ppm H

2

S

Accuracy

< 1 ppm H

2

S

Gas unit

1 ppm / 0.1 ppm

Signal update

every 1 second

Response time

< 30 seconds

Self-diagnostics

full functionality check at start-up

Power supply

11...30 VDC,

Power consumption

< 2 W

Analog outputs

2 × 4-20 mA / 0-10 V, user settable, freely configurable

Digital interface

RS485, Modbus RTU protocol
no galvanic isolation, common-mode voltage -7...+12 V,

Cable connections

screwless spring loaded terminals

Enclosure

grey ABS, wall mount, protection class IP65

Dimensions

H82 × W80 × D55 mm

Operating environment

industrial indoor and outdoor locations

Operating conditions

-30...+50 °C, 15...90 %RH

Maintenance interval

6 month

Sensor lifetime

2 years

CE marking

meets 2014/30/EU and EN61326-1 requirements

RS485 communication interface

Databits: 8

Parity: none
Stop bits: 1 or 2
Protocol: Modbus RTU

Supported Modbus functions:

03 - read multiple registers
06 - write single register

Communication parameters

Parameter

Permitted values

Default

Supported
baudrates

1200, 2400, 4800, 9600, 19200,
38400, 57600

9600

Data bits

8

8

Parity

none

none

Stop bits

1, 2

1

Protocol

Modbus RTU

Modbus functions

03 - read multiple registers
06 - write single register

Error codes

01 - illegal function
02 - illegal data address
03 - illegal data value
04 - slave device failure
(details of last error 04 can be read from register 0x0008)

Evikon MCI OÜ       

  Riia 185, Tartu 

[email protected]

Tel.: +372 733 6310        51014 Estonia 

 www.evikon.eu

Hydrogen sulfide transmitter

E2618-H2S

User manual