Evikon E2618-C2H4-24VAC User Manual Download Page 15

E2618-C2H4

Rev 10.05.2021

0x0005

5 / 40006 RW Baud rate *

1200, 2400, 4800, 9600, 19200,
38400, 57600

9600

0x0006

6 / 40007 RW Response delay, ms

1...255

10

0x0007

7 / 40008 RW Stop bits, parity bit *

1 – No parity bit, 1 stop bit
(default after factory reset)
2 – No parity bit, 2 stop bits
3 – Odd parity, 1 stop bit
4 – Even parity, 1 stop bit
NOTE: 3 and 4 are available
starting from the Software
version 0x218 (dec. 536)

1

0x0008

8 /40009

R

Last error code

1...255

-

0x0011 17 / 40018 RW

Technological:
age of last data in seconds
(read) / restart (write)

0...65535 s (read),
writing 42330 restarts the
instrument

-

0x00A5

165 /

40166

RW

Zero adjustments for gas
data, ADC

-32000...+32000 ADC units

0

0x00A6

166 /

40167 RW

Slope adjustment for gas

data

1...65535

512

0x00A7

167 /

40168 RW

Change rate limit for gas

data, units / s

1...32000, 0 - no limit

0

0x00A8

168 /

40169 RW

Integrating filter time

constant, s

1...32000 (seconds),

0 - no filter

0

Address Reg / MHR RW Description

Supported values (dec)

Default

0x00C9

201 /

40202

RW

Parameter tied to analog
output 1

0 – None
2 – Gas concentration
9 – Forced Modbus control,

value set in MHR / 40204

2

0x00CA

202 /

40203

RW

Parameter tied to analog
output 2

0 – None
2 – Gas concentration
9 – Forced Modbus control,

value set in MHR / 40205

2

0x00CB

203 /

40204

RW

Forced value for analog
output 1***

0...1000
(0.0%...100.0% of output scale)

0

0x00CC

204 /

40205

RW

Forced value for analog
output 2***

0...1000
(0.0%...100.0% of output scale)

0

15

Summary of Contents for E2618-C2H4-24VAC

Page 1: ...PluraSens Ethylene Transmitter E2618 C2H4 User Manual...

Page 2: ...6 Mounting dimensions 7 Sensor probe handling 8 Gas sensor replacement procedures 8 Electrical connections 9 Operation 11 Maintenance 11 Calibration 11 Delivery set 12 Order code for E2618 C2H4 option...

Page 3: ...Upper explosive limit UEL vol in air 36 0 Odor Faint sweet odor when pure Odor threshold 270 600 ppm so the odor is not an adequate warning property to prevent excessive exposure to ethylene Hazards...

Page 4: ...s Analog outputs 2 4 20 mA 0 10 V user settable Outputs assignment OUT1 2 gas OUT2 2 gas Output scale width 10 resolution Recommended 20 100 of the detection range Enclosure Grey ABS wall mount protec...

Page 5: ...m any maintenance operation with the power on Do not let water or foreign objects inside the device Removal of the PCB from the enclosure voids the warranty Do not touch the electronic components dire...

Page 6: ...the coverage area for a transmitter is usually between 2 5 to 5 meters radius For early leakage detection install the sensor as close as possible to the potential leakage sources flanges valves press...

Page 7: ...E2618 C2H4 Rev 10 05 2021 Mounting dimensions Wall mount version Duct mount version 7...

Page 8: ...the sensor head cap and remove the old filter Place a new filter into the cap and tighten it again NOTE Never stab or press the filter near its center where the sensor is located since this may damag...

Page 9: ...g the connection diagram below connect the analog outputs and digital interface terminals to the relevant devices according to your tasks PCB without PSU and relays Jumpers J1 OUT1 type open 4 20 mA c...

Page 10: ...the wire out and release the lever Use a twisted pair cable e g LiYY TP 2 2 0 5 mm2 or CAT 5 to connect the device to the RS485 network A and B on the device represent DATA and DATA respectively pola...

Page 11: ...pulses Normal operating measurement Continuously on or off Make sure that the transmitter is properly mounted the external devices connected power LED green on and control LED red constantly lit Place...

Page 12: ...ion Round rubber flange for Remote probe option Fixing clamp for Remote probe option Order code for E2618 C2H4 options E2618 options Order code Remote probe 3 m cable E2618 C2H4 RP33 3 Remote probe 10...

Page 13: ...g averaging filter Free assignment of each analog output to the chosen parameter Flexible setting of analog output scales for each output Output zero and slope adjustment for calibration The output sc...

Page 14: ...ple registers 06 Write a single register Error codes 01 Illegal function 02 Illegal data address 03 Illegal data value 04 Slave device failure details of last error 04 can be read from register 0x0008...

Page 15: ...Zero adjustments for gas data ADC 32000 32000 ADC units 0 0x00A6 166 40167 RW Slope adjustment for gas data 1 65535 512 0x00A7 167 40168 RW Change rate limit for gas data units s 1 32000 0 no limit 0...

Page 16: ...ll be ignored bit 6 0 1 current voltage output detected on output1 RO bit 7 0 1 current voltage output detected on output2 RO bit 8 0 1 LED deactivated activated bit 9 0 1 buzzer deactivated activated...

Page 17: ...n be used to assign a new ID to the instrument with an unknown ID When addressing by ID 0 the device shall be the only Modbus instrument in the network The device will not respond to the Master comman...

Page 18: ...Manufacturer will at its option either repair or replace a product that proves to be defective This warranty is void if the product has been operated in conditions outside ranges specified by the Manu...

Reviews: