E2618 series Modbus holding registers (continued)

Register addresses are shown 0-based, Addr in hexadecimal, Reg in decimal format.
Modbus holding register numbers MHR are shown in decimal 1-based format, and may be addressed either from 00001 or 40001 base.

0x00A2

162 / 40163 RW

Zero adjustment for temperature

-32000...+32000 (-320,00...+320,00 °C)

0

0x00A5

165 / 40166 RW

Zero adjustment for gas data, ADC

-32000...+32000 ADC units

0

0x00A6

166 / 40167 RW

Slope adjustment for gas data

1...65535

512

0x00A7

167 / 40168 RW

Change rate limit for gas data,

1...32000, 0 - no limit

0

0x00A8

168 / 40169 RW

Integrating filter time constant, s

1...32000 (seconds), 0 - no filter

0

0x00C9

201 / 40202

RW

Parameter tied to analog output 1

0-none
1- temperature
2- gas concentration

9- forced Modbus control, value set in MHR / 40204

2

0x00CA

202 / 40203

RW

Parameter tied to analog output 2

0-none
1- temperature
2- gas concentration

9- forced Modbus control, value set in MHR / 40205

2

0x00CB

203 / 40204

RW

Forced value for analog output 1***

0...1000 (0,0%...100,0% of output scale)

0

0x00CC

204 / 40205

RW

Forced value for analog output  2***

0...1000 (0,0%...100,0% of output scale)

0

0x00FF

255 / 40256

RW

Sensor and analog outputs status

bit[0]=0/1 - sensor present/absent, read-only!
bit[1]=0/1 - analog outputs deactivated/activated,
bit[2]= 0/1 - in case of sensor absent, turn signaling off/on analog output1,

bit[3]=0/1 - in case of sensor absent, turn on signaling with low current/high current
on analog output1; if bit[2]==0 this bit will be ignored,
bit[4]=0/1 - in case of sensor absent, turn signaling off/on analog output2

bit[5]=0/1 - in case of sensor absent, turn on signaling with low current/high current
on analog output2; if bit[4]==0 this bit will be ignored,
bit[6]=0/1 - current/voltage output detected on output1, read-only!

bit[7]=0/1 - current/voltage output detected on output2, read-only!
bit[8]=0/1 - LED deactivated/activated,
bit[9]=0/1 - buzzer deactivated/activated,

Note

 bits [8] and [9] are always 0 for E2618 series

62d

0x0100

256 / 40257

R

Raw temperature data, °C×100

signed integer, -4000...+8500 (-40,00...+85,00 °C)

-

0x0101

257 / 40258

R

Raw gas sensor data

ADC data 0...4095

-

0x0102

258 / 40259

R

Measured temperature, °C×100

signed integer, -4000...+12500 (-40,00...+125,00 °C)

-

0x0103

259 / 40260

R

Gas concentration, ppm / ‰

signed integer, -32000...+32000 (ppm / ‰)

-

0x0105

261 / 40262

RW

0% value for analog output 1

signed integer, -32000...+32000 (ppm / ‰)

0

0x0106

262 / 40263

RW

100% value for analog output 1

signed integer, -32000...+32000 (ppm / ‰)

1000

0x0107

263 / 40264

RW

0% value for analog output 2

signed integer, -32000...+32000 (ppm / ‰)

0

0x0108

264 / 40265

RW

100% value for analog output 2

signed integer, -32000...+32000 (ppm / ‰)

1000

* - The new value is applied after restart.
** - Broadcast slave ID 0 can be used to assign a new ID to device with unknown ID. When addressing by ID 0 the device shall be the
only Modbus instrument in the network. The device will not respond to Master command when addressed by ID 0.
*** - This value is dynamic and not kept in EEPROM after restart

Summary of Contents for E2618-C2H2

Page 1: ...series devices are suitable for a wide range of wires with cross section 0 2 1 5 mm2 The recommended wire stripping length is 8 9 mm Push the spring loaded terminal lever insert the wire end into terminal hole and release the lever Use twisted pair cable e g LiYY TP 2 2 0 5 mm2 or CAT 5 to connect the device to RS485 network Use one pair for A and B wires and the second pair for common 0 V and pow...

Page 2: ...s been an unauthorized modification E2618 C2H2_UM_EN Rev 04 05 2016 General technical data Sensor type long life metal oxide gas sensor Calibration acetylene C2H2 Gas unit ppm Sampling method diffusion Detection range 0 100 LEL 0 25 000 ppm 0 2 5 vol Resolution 0 1 LEL 25 ppm 0 0025 vol C2H2 Accuracy 5 LEL 125 ppm 0 125 vol C2H2 Warm up time 1 min Signal update every 1 second Response time 15 seco...

Page 3: ...30 to restart device no response will follow 0x0091 145 40146 RW Heater voltage pulse duration ms 0 1000 ms 0 0x0092 146 40147 RW Sensor voltage pulse delay ms 0 995 ms values 0 4 reserved to set sensor type 0 0x0093 147 40148 RW Sensor type specific parameter 0 65535 100 0x0094 148 40149 RW Sensor type specific parameter 0 65535 100 0x0095 149 40150 RW Sensor type specific parameter 32767 32767 3...

Page 4: ...r absent turn on signaling with low current high current on analog output1 if bit 2 0 this bit will be ignored bit 4 0 1 in case of sensor absent turn signaling off on analog output2 bit 5 0 1 in case of sensor absent turn on signaling with low current high current on analog output2 if bit 4 0 this bit will be ignored bit 6 0 1 current voltage output detected on output1 read only bit 7 0 1 current...