2/3 in. CCD
1/2 in. CCD
Field number 22
CCD reference: 1 in. camera 16.16 mm, 2/3 in. 11 mm, 1/2 in. 8.08 mm, 1/3 in. 6 mm, 1/4 in. 4 mm
Magnification on monitor =
*
Differs depending on the manufacturer.
Objective
magnification
Camera adapter
magnification
(0.63X)
Monitor diagonal length
*
CCD diagonal length
*
Adjusting the Microscope
Turn on the microscope light source and adjust the required points of
the microscope to make it ready for observation.
Set the light path of the UIS2 (UIS) trinocular tube to the camera light
path.
Adjusting the Camera and Monitor
Perform the adjustments such as color adjustment by referring to the
instruction manual of your camera and monitor.
} The center of the eyepiece image and that of the monitored image
may not coincide correctly. This is a function of the CCD adjustment
mechanism of the camera, and not a malfunction.
Adjusting the Perfocality between the Observed and Monitored
Images
} The parfocality adjustment makes re-focusing unnecessary when
the observed image is switched to the monitored image. The
parfocality adjustment requires the Allen screwdriver provided with
the microscope.
The parfocality adjustment range is ±1.5 mm. If the adjustment of
the adapter is not enough, also adjust the focusing feature of the
camera.
Using a high-power objective, look into the eyepieces and bring the
specimen into focus.
Switch to a low-power objective, set the camera light path and observe
the monitor image.
Using the Allen screwdriver provided with the microscope, loosen the
clamping screw
b
and
c
.
While observing the monitor image, adjust its focus by rotating the
parfocality adjustment area
a
under the camera mount.
When correct focusing is obtained, tighten the lock screw
b
and
c
using the Allen screwdriver.
Rotating the Camera
Loosen the straight photo tube clamping screw
c
.
Rotate the camera to the desired position and tighten the straight
photo tube clamping screw
c
firmly.
1
2
1
CAUTION
1
2
3
4
5
1
2
b
a
c
} The following diagram shows the areas of the imaging
field when eyepieces with a field number of 22 are used for
reference. The image field area is determined by the field of
view of the eyepieces and the size of the CCD seen through
the camera adapter.
A part of the image is cut off when a 1 in. CCD is used.
CAUTION
3
Operation
4
Imaging Field Area
EVIDENT CORPORATION
AX8224 02
Issued in April, 2022