EverPaw 5228P User Manual Download Page 25

23

                                                                                                                                                                                                                          

Problem

Possible Cause

Solution

Repeat the test after the 
meter and strip are placed in 
a warmer environment and 
allow the meter to warm up 
for a while before retesting.

Test again following the user 
guide for correct glucose 
measurement process. If you 
see HI again, please call your
 medical doctor for advice 
immediately. 

Repeat the test after the 
meter and strip are placed in 
a cooler environment and 
allow the meter to cool down 
for a period before retesting.

Please replace the two AAA 
batteries.

Please check the test strip to 
see if it is damaged or used.   
In either case, please discard 
the strip and repeat the test 
using a new strip.

Repeat the test with a new 
test strip. If E_6 continues 
to show up on screen, please 
call a customer service 
representative.

The meter can perform the 
glucose measuring without 
storing the test results into 
the meter's memory.  Please 
write down the test results 
into the data logger before 
you call a customer service 
representative.

Test again following the user
 guide for correct glucose 
measurement process. If you 
see LO again, please call your 
medical doctor for advice 
immediately. 

The meter is operating
in an ambient 
temperature below 
10 

 or 50

 which is

the lowest range of 
operating temperature.

(1) 
The meter displays 
an E_1 error 
message.                     

(2) 
The meter displays 
an E_2 error 
message.   
                  

The test result is 
lower than the 
measuring range 
of 1.1 mmol/L

 (20 mg/dL).

(3) The meter displays 
an E_3 error message. 

                    
(5)
The meter displays
 an E_5 error 
message.                     

          

(6)
The meter displays
an E_6 error 
message.         

(8) 
The meter displays
HI on screen.

(9) 
The meter displays
LO on the screen.

The meter is operating 

in an ambient 

temperature above 

40 

 or 104

 which is

 the cap of operating 

temperature.

The battery is too low 
to operate the meter.

1.If you apply the 
sample before blood 
drop icon appears on 
the screen.
2.The inserted test strip 
has been wet or used.

The meter is not 
working properly, 
either because of a 
defective meter or 
a defective strip.

The memory chip 
of the meter could 
be damaged or 
malfunctioning.

The test result is 
higher than the 
measuring range 
of 33.3 mmol/L

 (600 mg/dL).

(4)
The meter displays 
an E_4 error 
message.     

Please re-test by inserting a 
new test strip ensuring an 
adequate amount of blood 
comes into contact with the 
test strip.

The blood sample is 
not sufficient .

(7)
The meter displays
an E_7 error 
message.  

Summary of Contents for 5228P

Page 1: ......

Page 2: ...structions for use Keep away from sunlight Do not re use Use by date Batch code In vitro diagnostic use in animals Temperature limit Caution refer to accompanying documents Please refer to safety rela...

Page 3: ...n 8 Coding the Meter 9 Running a Control Solution Test Blood Testing 10 Running a Blood Glucose Test 11 Recalling the Memory and Viewing theAverage 12 Expected Values 13 Transmitting Results Maintenan...

Page 4: ...homeasanaidinmonitoringtheeffectiveness ofadiabetescontrolprogram TheEverPaw model 5228P BloodGlucoseMonitoring Systemisnotintendedforthediagnosisoforscreeningfor diabeticpets ContentsoftheSystem This...

Page 5: ...PawControl Solution 6 Sterile Lancets 7 Lancing Device 8 Package Inserts for Test Strips and Lancing Device Asingle level control solution is provided in the Starter Kit You may purchase any control s...

Page 6: ...r will automatically turn on 2 LCD Display Guide you through the test using symbols and simple messages 3 M Key Power ON OFF also for memory recalling mode please refer to manual for detailed function...

Page 7: ...Please keep the test strip vial away from children or pets The test strips and vial cap can present a choking hazard Please be cautious when removing the lancet Take the lancet out carefully Always p...

Page 8: ...rt a new test strip into the meter while the meter is off The meter will turn on automatically Ablinking blood icon will be displayed on the screen 2 Choose a capillary sampling site 3 Apply the blood...

Page 9: ...n the display please follow the above procedure to replace the batteries again 5 Please dispose of batteries according to your local ordinances 1 Turn off the meter by pressing M button and hold for 2...

Page 10: ...flashes 5 Indicates the battery status 6 Indicates current displayed result is an average 7 Appears when alarm is on 8 Month 9 Day 10 Hour 11 Minute 12 Dog Blood Testing 13 Cat Blood Testing 14 Contro...

Page 11: ...2030 After the year setting is confirmed you will see the month segment flashing on the screen Set the Month Press C key to adjust the month until the desired month is displayed and then press the M k...

Page 12: ...rm clock in the lower left corner with a time segment display it means that you are in the alarm setting mode If you see only AL without the icon of the small alarm clock it means that you are not in...

Page 13: ...uring the 7 14 28 60 and 90 calendar days preceding the current date and time settings Delete Memory Press C key to toggle between a flashing dEL for memory deletion or a non flashing dEL If you want...

Page 14: ...strip the meter will turn on automatically and display the last code number saved If the code number on the screen matches the code number printed on the vial label proceed with the test If the code...

Page 15: ...een indicating the control test mode The test result will not be stored in the memory 3 Squeeze a small amount of Control Solution on a flat surface and apply to the front edge of the test strip 4 Wai...

Page 16: ...and High level Control Solution are intended to check the monitoring system in different measurement ranges 6 Shake the vial discard the first drop of control solution and wipe off the dispenser tip...

Page 17: ...dy of the lancing device until you hear the click When the cap is properly closed in locked position you should see the arrow and release button are aligned see Figure 3 4 Holding the cap and select l...

Page 18: ...ter while the meter is off The meter will turn on automatically Ablinking blood icon will be displayed on the screen If no action is taken in 2 minutes the meter will turn off automatically 8 Use the...

Page 19: ...eter should be removed and the meter should be disposed in accordance with local ordinances 9 Apply a small blood drop to the front edge of the test strip and the blood should be pulled into the confi...

Page 20: ...f press M key to turn on the meter After a brief full screen display the meter will display the total number of test data stored in memory Then the screen will display 001 as the sequential number of...

Page 21: ...ward from 7 day average to 90 day average The average display will move from 90 days backward to 7 days if you continue to press C key 3 Press the M key and hold for only 2 seconds to exit the recall...

Page 22: ...ill vary from time to time depending on food intake medication dosages health stress or exercise Consult your veterinarians for the target glucose value appropriate for your pets EverPawmeter gives pl...

Page 23: ...soft cloth that has been slightly dampened with alcohol 3 Please handle with care and do not drop the meter 4 If this meter is used by veterinarians infection control policies should be strictly foll...

Page 24: ...ple is not sufficient HI The glucose level is too high LO The glucose level is too low Following is a summary of all Error Messages These messages help to identify certain problems but do not appear i...

Page 25: ...lease call your medical doctor for advice immediately The meter is operating in an ambient temperature below 10 or 50 which is the lowest range of operating temperature 1 The meter displays an E_1 err...

Page 26: ...oagulation of fresh capillary or venous whole blood 2 The test strips may be used at altitudes up to 10 744 feet 3 275 m without an effect on test results 3 Pets suffering from severe dehydration shou...

Page 27: ...emory 500 measurement results with date and time Auto power off after 2 minutes without action when a strip is inserted or 60 seconds if there is no strip inserted Traceability of glucose monitoring s...

Page 28: ...device failure due to owner misuse or negligence or normal wear and tear If you have a question about your EverPaw model 5228P Glucose Meter or this warranty please contact TM EverPaw LLC Brooklyn NY...

Page 29: ......

Reviews: