16
Guidance and manufacturer’s declaration - Electromagnetic immunity - Enclosure port
Phenomenon
Basic EMC
standard or test
method
Professional
healthcare facility
environment
Immunity test levels
HOME HEALTHCARE
ENVIRONMENT
ELECTROSTATIC
DISCHARGE
IEC 61000-4-2
± 8kV contact
± 2 kV, ±4kV ±, ±8 kV, ±15 kV air
Radiated RF EM fields
IEC 61000-4-3
a)
10 V/m b) 80MHz - 2.7 GHz
80% AM at 1kHz
Proximity fields
from RF wireless
communications
equipment
IEC 61000-4-3
COMPLIANT
NOTE: Further information about distances to
be maintained between portable and mobile RF
communications equipment (transmitters) and
the 60321000/NT6/NT66 can be requested from
the manufacturer using the contact information
provided in this manual. However, it is advisable
to keep the electromechanical aerosol equipment
at an adequate distance of, at least, 0.5 m from
mobile phones or other RF communications
transmitters to minimize possible interference.
RATED power
frequency magnetic
fields.
IEC 61000-4-8
30 A/m c)
50 Hz or 60 Hz
a) The equipment is suitable for use in Home Health Environments and Professional Health Care
Environments limited to patient rooms and respiratory treatment facilities in hospital or clinics. The
more restrictive IMMUNITY acceptance limits have been considered and applied.
b) Before modulation is applied.
c) This test level assumes a minimum distance of at least 15 cm between the ME EQUIPMENT or ME
SYSTEM and sources of power frequency magnetic fields.