EUTECH INSTRUMENTS WATERPROOF PHTESTR 1 PH TESTER Instruction Sheet Download Page 3

Eutech Instruments

Oakton Instruments

www.eutechinst.com

www.4oakton.com

Lanyard
Connection

Replaceable

Sensor

S P E C I F I C A T I O N S

 Model

pHTestr  1,  2

pH Range

-1.0 to 15.0 pH

Resolution

0.1 pH

Accuracy

pHTestr 1 : ± 0.2 pH
pHTestr 2 : ± 0.1 pH

ATC

Yes (Except pHTestr  1)

Calibration

pHTestr 1 :

1-point (4.0, 7.0 or 10.0)
pHTestr 2 :

3-point (4.0, 7.0 & 10.0)

Auto Power- Yes
Off/Hold
Auto-Buffer

Yes (pH 4, 7, & 10)

Recognition
Battery Life

3 x 1.4V (Type: A 76,
LR44, V13GA) ; >60 hours

Dimension/ Tester : 15 x 3.8 cm ; 90g
Weight

Boxed : 18.5 x 7 x 5 cm ; 170g

Made in Malaysia

Twist-off

Battery

Compartment

Amplified

Electrode

68X247809 2/05 Rev 3

ELECTRODE REPLACEMENT:

 One of the

benefits in selecting the Waterproof pH
tester is that you can now replace the
electrode whenever needed keeping the
body of the tester intact. When the tester
fails to calibrate, gives fluctuating readings
in buffers, or shows error messages ‘E2’
or ‘OR’ in a buffer, and the procedures in
the maintenance section do not help,
you need to change the electrode (see
pictures for instructions).• With dry hands
firmly grip the ribbed collar (Picture A)
and slowly twist the body counter
clockwise until it is completely removed.
Save both the electrode module collar
and O-ring inside for later use. Pull the
old electrode module straight out from
the bottom of the tester • Align the four
tabs on the new module so they match
the four slots on the tester (Picture B).
Note: Older testers may have only two
slots. In this case, break off the two small
tabs using a needle-nose pliers • Gently
push the module onto the bottom end to
fully seat it in position. Put the small O-
ring all the way into the ribbed collar.
Then push the collar on over the module
and hand-tighten by firmly twisting it
clockwise.

Rotate collar away from you

Rubber

Gripper

Band

Ribbed Collar

Small O-Ring

(Hidden)

Large O-Ring

Replacement

Electrode

Insert Electrode Small Tabs

REMOVE THE OLD ELECTRODE

REPLACE THE ELECTRODE

Large Tab

 (Picture B)

 (Picture A)