background image

  CoND 

Setting up the 
Conductivity 
Parameter 

 

1. 

From            press 

ENT

.  

PArA 

 
Then use 

c

 and 

d

 keys to 

select  

  CoND 

 
 

2. 

Press 

ENT

 

3.

 Then use 

c

 and 

d

  keys to 

select either  
 
 

  OR  

 

13

 

  YES  

A

.

CAL 

  NO  

A

.

CAL 

Selecting 

YES

 means the 

automatic calibration will be 
activated. Press 

ENT

 to confirm. 

 
4. 

Then use 

c

 and 

d

 keys to 

select either  
 
 

   OR  

 
 

  YES  

SPC 

  NO  

SPC 

Selecting 

YES

 means the 

Single Point Calibration will be 
activated. Press 

ENT

 to confirm 

and return to the main menu. 
Selecting 

NO 

means multipoint 

calibration will be activated. 

 
 
 
 

Summary of Contents for PCSTestr35

Page 1: ...QUICK GUIDE MULTI PARAMETER TESTER 35 pH Conductivity TDS Salinity Temperature PCSTestr35 PCTestr35 PTTestr35 Technology Made Easy Part of Thermo Fisher Scientific 68X441601 Rev 0 Jan 08...

Page 2: ...ecute before using your meter STAGE 1 Conditioning and temperature calibration STAGE 2 Setup the system and the parameters STAGE 3 Calibrate the parameters BEFORE YOU BEGIN you must first load batteri...

Page 3: ...done once in a long while as the calibration for temperature will remain accurate unless used under severe conditions FOLLOW THESE STEPS 1 Press the ON key The tester is now in Measurement mode 2 Pla...

Page 4: ...e according to the reference thermometer 7 Press ENT to confirm and return to the PH measuring mode NOTE Once calibrated you don t have to do this for any other parameter as temperature is common to a...

Page 5: ...odule Electrode Module Large O Ring Small O Ring Small Tab Large Tab STAGE 2 SET UP 1 Turn your tester off Then hold the ENT key down and press ON OFF to enter the Setup mode 2 The display will show S...

Page 6: ...manual changes have been done during set up you are required to save the setting 1 Press both c and d keys at the same time and release 2 Press again until you see SA YES 3 Use either c OR d key to t...

Page 7: ...select UNit rST Use this to reset the calibration or factory settings Use this to turn on or off the AUTO OFF function SET A off Use this to set the temperature coefficient SET TC Set ATC Use this to...

Page 8: ...e pH parameter select by pressing ENT Then select by pressing ENT to reset your previous settings OR use the c and d keys to select to reset the factory settings Press ENT to confirm and return to 4 T...

Page 9: ...d Set a off 2 Press ENT to select it Use the c and d keys to switch between 9 and NO a off YES a off Selecting YES means your tester will turn OFF automatically if it is not used after a 8 5 minutes S...

Page 10: ...t then press ENT when done Turning on or off the Automatic Temperature Compensation 1 Use the c and d keys from and find 2 Press ENT to select it 3 Use the c and d keys to select either OR Then press...

Page 11: ...T and find Set C F 2 Press ENT to select it 3 Use the c and d keys to select either OR 11 C SET Press ENT when done F SET 4 Save the changes as described on page 6 END OF SYSTEM SETUP PARAMETER SETUP...

Page 12: ...you want to set up 12 Setting up the pH Buffer PH 1 At the above display press ENT 2 Use the c and d keys to select either OR Nist BUFF USA BUFF Then press ENT to confirm 3 Again use the c and d keys...

Page 13: ...ther OR 13 YES A CAL NO A CAL Selecting YES means the automatic calibration will be activated Press ENT to confirm 4 Then use c and d keys to select either OR YES SPC NO SPC Selecting YES means the Si...

Page 14: ...c and d keys to select TDS 2 Press ENT 3 Use the c and d keys to change the TDS value For example 0 67 FACT 4 Press ENT to confirm and return to main menu Setting up the Salinity parameter SALT 1 From...

Page 15: ...to the main menu 5 Save the changes as described on page 6 END OF PARAMETER SETUP STAGE 3 CALIBRATING THE PARAMETERS Calibrating PH 1 In the measurement mode at press the CAL key PH 2 Place the teste...

Page 16: ...rm Calibrating COND COND 1 At the display press the CAL key NOTE In your set up for COND the tester must be in AUTO CAL mode NOT manual CAL 2 Place the tester in the buffer solution 3 Follow the same...

Page 17: ...d keys to scroll to the calibration standard value and press ENT The tester will return to measurement mode Calibrating SALT 1 At the display press the CAL key 2 Place the tester in the buffer soluti...

Page 18: ...D key is pressed again This function is useful when taking measurements in a dim environment It allows you to hold the reading until you are able to move to a well lit area MEASUREMENT 1 To start meas...

Page 19: ...return the tester freight prepaid and correction will be made without charge Out of warranty products will be repaired on a charge basis Return of Items Authorization must be obtained from your distri...

Page 20: ...isted below Oakton Instruments P O Box 5136 Vernon Hills IL 60061 USA Tel 1 888 462 5866 Fax 1 847 247 2984 E mail info 4oakton com Web site www 4oakton com Eutech Instruments Pte Ltd Blk 55 Ayer Raja...

Reviews: