EUTECH INSTRUMENTS Ion 700 Instruction Manual Download Page 15

 

12 

 

PROBLEM CAUSE 

SOLUTION 

From any screen 

 

Invalid key;  
Button does not  
work in the current 
operation mode. 

Press alternate key.  
Select valid buttons 
depending on mode. 

Ion CAL mode 

Ion slope calibration error; 
calibration solutions are  
outside acceptable range 
(15 to 90 mV per decade). 

Recalibrate using fresh ion 
solutions. Use required ISA. 
Replace electrode. 

Ion CAL mode 

Calibration points  
are more than 1  
decade apart. 

Recalibrate using 
consecutive calibration 
values 1 decade apart  
(ex. 1.0 and 10.0). 

Ion CAL mode 

Calibration attempted 
with only 1-point during 
ion calibration. 

Recalibrate using 2 or more 
ion calibration points. 

pH CAL mode 

Calibration attempted 
using only 1-point that 
was not pH 7 (USA) or 
pH 6.86 (NIST). 

Recalibrate using 2 or more 
pH points or use pH 7 (USA) 
or use pH 6.86 (NIST). 

Err with electrode 
icon 

 

Calibration error.  
Buffer value does not 
match value displayed or 
electrode is disconnected 
or failing. 

Use fresh buffer solutions. 
Check electrode connection. 
Clean & recondition 
electrode. 
Replace electrode. 

Ur (Under range) 

Or (Over range) 

 

Measured value is out  
of range.  

Electrodes not connected. 

Electrode clogged, dirty  
or broken. 

Meter not calibrated.  

Wrong temp value. 

Make sure electrode is 
connected.  

Clean or replace electrode. 

Recalibrate the meter. 

Treat samples to bring 
within meter measuring 
range.  
Reset meter.  

Summary of Contents for Ion 700

Page 1: ...Technology Made Easy Instruction Manual pH 700 pH mV C F Bench Meter Ion 700 pH mV Ion C F Bench Meter Part of Thermo Fisher Scientific 68X541701 Rev 3 Nov 2010...

Page 2: ......

Page 3: ...Ion 700 Only 6 4 STORING AND RECALLING DATA 7 5 SETUP FUNCTIONS 8 5 1 P1 0 CAL Calibration 8 5 2 P2 0 ELE Electrode Information 8 5 3 P3 0 ConF Configuration 9 5 4 P3 1 rdY Ready Stability Indicator 9...

Page 4: ...ese parameters the Ion 700 features direct ion concentration measurement of various ions mono and divalent Each meter includes a convenient slide out card for quick reference Some configurations inclu...

Page 5: ...to Temp during pH calibration mode Press and hold for 5 seconds to enter SETUP mode Toggles between measurement and calibration modes In SETUP mode returns user to the measurement mode MI Memory Inser...

Page 6: ...n for pH ORP Redox or Ion Selective Electrodes ISE REF Pin connection for half cell reference electrodes Requires separate half cell BNC electrode Note REF is not commonly used and is not required ATC...

Page 7: ...tion ATC probe for best accuracy Without temperature compensation pH accuracy will worsen as samples deviate from 25 C and pH 7 1 Press as needed to select pH 2 Dip the pH and ATC electrodes into pH b...

Page 8: ...e compensation and measurement is both accurate and stable so require frequent calibration isn t required Temperature calibration is recommended upon electrode replacement whenever the temperature rea...

Page 9: ...odidetriiodide and stir 3 Press when the reading is stable The primary display shows the relative millivolt value R mV while the secondary display shows the factory default mV value 4 Adjust the R mV...

Page 10: ...pe value will be displayed briefly if the calibration is successful SLP Err indicates that the calibration for the current point was not successful This occurs when the slope mV difference between two...

Page 11: ...tomize your instrument operation During measurement press and hold for 5 seconds to enter SETUP mode Press or to change programs or change options Press to select the program or confirm selection Pres...

Page 12: ...r that corresponds to pH values above 7 00 If the measured pH value was 5 23 when setup mode was entered the pH slope that corresponds to values lower than 7 00 will be available to view If no calibra...

Page 13: ...number of Calibration Points pH and Ion only Press to confirm 5 7 P4 0 rSt Reset Press Press or to select Yes or No If Yes press or to select Cal calibration reset only or FCt complete reset to facto...

Page 14: ...electrolyte b Broken electrode b Replace electrode c External noises or induction e g electrical noise caused by a nearby motor c Remove or switch off interfering device Unstable reading d Dirty elect...

Page 15: ...calibration Recalibrate using 2 or more ion calibration points pH CAL mode Calibration attempted using only 1 point that was not pH 7 USA or pH 6 86 NIST Recalibrate using 2 or more pH points or use p...

Page 16: ...most common electrodes available Ammonia NH3 Iodide I Ammonium NH4 Lead Pb 2 Bromide Br Lithium Li Cadmium Cd 2 Nitrate NO3 Calcium Ca 2 Nitrogen Oxide NOx Carbon Dioxide CO2 Perchlorate CIO4 Chlorid...

Page 17: ...p to 150 mV Temperature Range 0 0 to 100 0 C 32 0 to 212 0 F Resolution 0 1 C F Accuracy 0 3 C 0 5 F 0 to 70 C Compensation Automatic or Manual 0 to 100 C 32 to 212 F Calibration Offset in 0 1 increme...

Page 18: ...nce by customer Unauthorized modification or misuse Operation outside of the environment specifications of the products 10 RETURN OF ITEMS Authorization must be obtained from our Customer Service Depa...

Page 19: ......

Page 20: ...ts 625 E Bunker Court Vernon Hills IL 60061 USA Tel 1 888 462 5866 Fax 1 847 247 2984 info 4oakton com www 4oakton com Eutech Instruments Pte Ltd Blk 55 Ayer Rajah Crescent 04 16 24 Singapore 139949 T...

Reviews: