Instruction Manual  

 

                       CyberScan DO 1500 

 

 

16

 

3. Press 

enter

 to accept the current value or press

 

 

or 

 

to 

adjust and set the value to be 

1

 or 

2

 

4. Press 

enter

 key to confirm selection and to go to next setup 

page OR press 

std

 to exit from this page without confirming the 

change mode. 

 

 
 

 

 

Summary of Contents for CYBERCOMM PRO FOR CYBERSCAN DO 1500

Page 1: ...Instruction Manual CyberScan DO 1500 Dissolved Oxygen Meter 68X292336 Rev 1 01 04 Technology Made Easy...

Page 2: ...yberScan DO 1500 bench meter as possible If there are doubts in the use of the meter please do not hesitate to contact the nearest Eutech Instruments Authorised Distributors or call us at 65 6778 6876...

Page 3: ...r 21 4 1 8 Setup P8 0 Baud Rate 23 4 1 9 Setup P9 0 Parity Bit 25 4 1 10 Setup P10 0 Stop Bit 27 4 1 11 Setup P11 0 Print Data 29 4 1 12 P12 0 Clear Stored Data 31 4 1 13 P13 0 Clear Buffer Values 33...

Page 4: ...laying easy to understand prompts This meter is designed to provide all the information necessary to guide the user through the process of measuring dissolved oxygen with a series of prompts on the sc...

Page 5: ...Connectors 1 Review the layout and arrangement of the rear connector panel 2 Connect the power adapter s output power jack to the meter s rear panel DC input power socket and plug in the adapter to a...

Page 6: ...peration The 5010 probe need be plugged into the meter only as the meter supplies both its operation power and stirring power 1 Prepare the electrode as described in its instruction manual 2 Plug the...

Page 7: ...s std key at the Standardize mode to exit to Measurement mode without confirming the calibration OR to exit from Setup mode Press mode to select operation mode to be mg L or sat Press setup key to acc...

Page 8: ...Instruction Manual CyberScan DO 1500 5 3 3 Screen Display Familiarise yourself with the layout of the digital screen display...

Page 9: ...erating parameters While in the setup mode you may Press the std key at any time to return to the Measure screen without making a change or selection in the setup mode Note new change will not be effe...

Page 10: ...P 8 0 Set the Baud Rate Set baud rate as 4800 9600 19200 or 38400 bps P 2 0 Set the pressure 425 850mmHg Adjust and select Pressure value Print mmHg P 9 0 Set the Parity Bit Set parity bit as none 0...

Page 11: ...etup P 1 0 View the Calibration Data This setup menu allows you to view the last calibrated Cal factor in Sat and mg L mode press to accept enter press to set value View 1 Access the Setup pages while...

Page 12: ...Instruction Manual CyberScan DO 1500 9 2 Press enter to access View Last calibrated Cal factor setup page press to accept enter...

Page 13: ...to exit from this page without confirming the change mode If you access the setup page from Sat mode and if you have calibrated one point 100 Sat then you can view the last calibrated Cal Factor Shoul...

Page 14: ...set the atmospheric pressure mmHg value You are able to set the salinity value in the range of 450mmHg to 825 mmHg press to accept enter press to set value mmHg To set Pressure 1 Access the Setup page...

Page 15: ...and keys 5 Press enter key to confirm selection and to go to next setup page OR press std to exit from this page without confirming the change mode press to accept enter press to set value mmHg You ca...

Page 16: ...you to set the Salinity as ppt value You are able to set the salinity value in the range of 0 0 45 0 ppt press to accept enter press to set value Salinity To set Salinity 1 Access the Setup pages whi...

Page 17: ...y adjustment setup page as shown below 4 Adjust the salinity value using the and keys 5 Press enter key to confirm selection and to go to next setup page OR press std to exit from this page without co...

Page 18: ...the number of user cal points for Sat calibration to be either 1 or 2 press to set value press to accept enter Sat To Select Number of User Cal Points 1 Access the Setup pages by pressing setup while...

Page 19: ...16 3 Press enter to accept the current value or press or to adjust and set the value to be 1 or 2 4 Press enter key to confirm selection and to go to next setup page OR press std to exit from this pag...

Page 20: ...s you to enable the auto calibration for Sat and mg L modes press to set value press to accept enter Sat To Enable Disable Auto Cal 1 Access the Setup pages by pressing setup while in measurement mode...

Page 21: ...uction Manual CyberScan DO 1500 18 3 Press or to select YES or NO 4 Press enter key to confirm selection and to go to next setup page OR press std to exit from this page without confirming the change...

Page 22: ...pt enter This setup option allows you to select unit of measure for Temperature either in C or F To Select Temperature Unit 1 Access the Setup pages menu from the measurement mode by pressing the setu...

Page 23: ...Manual CyberScan DO 1500 20 3 Press or key to select C or F 4 Press enter key to confirm selection and to go to next setup page OR press std to exit from this page without confirming the change mode s...

Page 24: ...isplayed on the screen whenever reading has stabilised thus minimises guesswork ST AB L E press to set value press to accept enter To Set Stability Indicator 1 Access the Setup pages menu from the mea...

Page 25: ...tion Manual CyberScan DO 1500 22 3 Press or key to select ON or OFF 4 Press enter key to confirm selection and to go to next setup page OR press std to exit from this page without confirming the chang...

Page 26: ...te bits per second of the communication protocol interface P rint press to set value press to accept enter To Set Baud Rate 1 Access the Setup pages menu from the measurement mode by pressing the setu...

Page 27: ...4800 9600 19200 or 38400 4 Press enter key to confirm selection and to go to next setup page OR press std to exit from this page without confirming the change mode save view pres s to s et value pres...

Page 28: ...cation protocol interface pres s to s et value pres s to accept enter P rint press to set value press to accept enter Print To Set Parity Bit 1 Access the Setup pages menu from the measurement mode by...

Page 29: ...3 Press or key to toggle between toggle between 0 none 1 odd or 2 even 4 Press enter key to confirm selection and to go to next setup page OR press std to exit from this page without confirming the ch...

Page 30: ...ication protocol interface pres s to s et value pres s to accept enter P rint press to set value press to accept enter Print To Set Stop Bit 1 Access the Setup pages menu from the measurement mode by...

Page 31: ...CyberScan DO 1500 28 3 Press or key to toggle between 1 or 2 4 Press enter key to confirm selection and to go to next setup page OR press std to exit from this page without confirming the change mode...

Page 32: ...e port Note all the communication protocol for both the meter and computer printer must match before successful printing can be performed press to set value press to accept enter Print To Select Print...

Page 33: ...an DO 1500 30 3 Press or key to toggle between Curr or Store 4 Press enter key to confirm selection and to go to next setup page OR press std to exit from this page without confirming the change mode...

Page 34: ...ta sets will be overwritten by any new data sets in the event when the stored locations have exceeded pres s to s et value pres s to accept enter P rint press to set value press to accept enter clear...

Page 35: ...nter key to confirm selection and to go to next setup page OR press std to exit from this page without confirming the change mode When you press enter key to confirm the clearing of memory meter displ...

Page 36: ...he standardised buffer values press to set value press to accept enter clear BUFFER Sat mg L To Clear Buffer Values 1 Access the Setup pages menu from the measurement mode by pressing the setup key 2...

Page 37: ...to confirm selection and to go to next setup page OR press std to exit from this page without confirming the change mode If you are doing user reset from sat setup page it will clear both sat and mg L...

Page 38: ...ion Accepted window for 100 sat 50 200 Sat 1 Press std to enter to Standardize screen Upper display shows the present measured value while lower display shows the pressure compensated 100 sat 2 Hold p...

Page 39: ...shows the 0 0 Sat 2 Dip the DO electrode into 0 Sat solution 3 Wait for the upper display value to stabilise Press enter key to confirm calibration Meter blinks the cal values for few seconds and pro...

Page 40: ...ter to Standardize screen Both Upper and lower display shows the present measured value 2 Dip the electrode into the standard solution Wait for the upper display value is stabilised Use keys to adjust...

Page 41: ...present measured value while lower display shows theoretical value Based on pressure and salinity setting 2 User can just confirm the theoretical reading 3 Acceptance window is 70 of the theoretical...

Page 42: ...asured value 2 Use the keys to adjust the known standard value 3 Window provided for the adjustment is 70 of the present reading Lowest value that can be set is 2 00mg L highest value is 60 00 mg L 4...

Page 43: ...ATC temperature offset 1 From the measurement mode press std and followed by mode key to go into the temperature calibration mode 2 Use the keys to adjust the offset 3 After set the value press enter...

Page 44: ...ves data from the meter s memory on the Last In First Out LIFO basis The screen displays the last stored memory location To view stored data in that particular memory location press enter key If you w...

Page 45: ...figuration in terms of baud rate parity bit and stop bit Please check with the printer s or peripheral s manufacturers for details of any specific settings of the device in use If there is an error me...

Page 46: ...gen and therefore more current The partial pressure that oxygen exerts is fixed for a given temperature and atmospheric pressure This fixed level corresponds to the saturation level for a given temper...

Page 47: ...y 0 5 Full Scale 1 LSD Temperature range 0 0 to 45 0 C 32 0 to 113 0 F Barometric pressure compensation 450 to 825 mmHg Salinity Correction 0 to 45 ppt Memory 100 data sets Display 105 x 75 MM SCREEN...

Page 48: ...damp cloth from time to time 11 TROUBLESHOOTING The CyberScan DO 1500 displays pertinent error messages to guide you should an error occur with a measurement or meter operation Message Description Er...

Page 49: ...e EC 620 SSP Data Acquisition Software and 220VAC power adapter 2 round pin EURO type center negative EC 637 DOM 6 membrane caps polishing disk and electrolyte solution 60X030115 110 120 VAC power ada...

Page 50: ...y customer Unauthorised modification or misuse Operation outside of the environmental specifications of the products Return of Items Authorisation must be obtained from your Eutech Instruments Authori...

Page 51: ...equipment in a residential area may cause interference in which case the user at his own expense will be required to take whenever measures may be required to correct the interference This product is...

Page 52: ...t Eutech Instruments distributor or visit our website listed below Manufactured by Eutech Instruments Pte Ltd Blk 55 Ayer Rajah Crescent 04 16 24 Singapore 139949 Tel 65 6778 6876 Fax 65 6773 0836 E m...

Reviews: