8.7
Viewing TDS Calibration Data – P5.0
Program 5 is a “view only” option, which displays the value at which calibration was
performed.
P5.1: view Calibrated TDS value (if no calibration performed for this range,
controller displays ‘- - - - ‘).
These parameters will change each time you recalibrate the controller.
P5.0
C L
P5.1
CAL
300
Press
▲
and
▼
keys together (ESCAPE) at anytime, to leave the Setup mode and
return to Measurement mode.
P5.0: Viewing Calibrated TDS data
1.
Key in the password “022” as per procedure in Section 5.
2.
Press
▲
key until screen displays P5.0 and CAL. Press ENTER.
3.
Screen will scroll P5.1, ‘CAL’, then Calibrated value (in ppm or ppt).
4.
Press
▲
and
▼
keys together (ESCAPE) twice, to return to measurement
mode.