7.4.3 Reverting to factory default settings
Use this parameter to reset all settings to factory default. Changing from “no”
to “YES” and pressing the ENTER key resets all settings to factory default.
WARNING: If “Yes” is selected, all settings input
will be overwritten!
1) Follow directions in 7.4.1 to enter Configuration
mode.
2)
Press the ENTER key
. Scroll with the
∆∆∆∆
or
∇
∇
∇
∇
keys until the upper
display shows “no” or” YES”, and the lower display shows “deF” (default).
3)
Press the
∆∆∆∆
or
∇
∇
∇
∇
keys
to select no or yes. Selecting “YES” and pressing
the ENTER key will overwrite all user-settings and automatically return to
Measurement mode.
4)
Press the ENTER key
to confirm your selection and return to Advanced
Set-up mode, or return to Measurement mode by pressing the
∆∆∆∆
and
∇
∇
∇
∇
keys
simultaneously (escape).
7.5 Calibration (CAL) sub-function
The calibration procedure in Advanced Set-up mode is identical to the
procedure in the Calibration mode. The only difference is that the controller
will revert back to Set-up mode (instead of Measurement mode) after
calibration is completed.
7.5.1 Entering Calibration mode from Advanced Set-up mode
1)
Enter Advanced Set-up mode
. Press the ENTER key.
2)
Press the
∆∆∆∆
or
∇
∇
∇
∇
keys
to scroll until the upper display shows “CAL”.
3) See section 6 for complete calibration procedures.
no
dEF
SETUP
HOLD