12
SETUP
HOLD
5.2 Output Range Setting
Output of pH 500 transmitter is via the 2-wire power supply
loop. Both the 4 mA and 20 mA output can be assigned to
specific pH or ORP values, for a more refined output.
1. Press ENT key and use
▲
or
▼
keys to scroll till LCD
displays
‘
r n g
’. Press ENT again. Upper display shows pH value,
‘0.00’ (or – 1000 mV) while lower display shows ‘
r 4.0
’.
2. Assign pH or ORP value to 4 mA output current. Use
▲
or
▼
keys to change pHor ORP value. Press ENT to accept
value.
3. Upper display shows pH value,’
14.00
’ (or 1000 mV) while
lower display shows ‘
r 20.0
’.
4. Assign pH or ORP value to 20 mA output current. Use
▲
or
▼
keys to change
pH or ORP value. Press ENT to accept value.
5. Press
both
▲
or
▼
keys together to return to ‘ MEAS ‘ mode.