Instruction Manual alpha-pH 1000
25
7.6
Configuration (ConF) sub-function
This group of parameters lets you configure the controller to suit your requirements.
7.6.1
Entering the Configuration sub-function
1)
Enter Advanced Set-up mode
. Push the ENTER key and scroll to Advanced Set-up security code “22”.
Push the ENTER key again.
2)
Press the
▲
or
▼
key
to scroll until the upper display shows “ConF”.
7.6.2
Selecting pH, ORP% or ORP mV measurement
This parameter group lets you select pH, mV, or % readings, and input type symmetrical or asymmetrical.
Symmetrical mode is used in conjunction with a potential matching pin – in an electrically noisy environment.
Please refer to section 3, page 6 for a graphical representation.
1)
Follow directions in 7.6.1
to enter Configuration mode.
2)
Press the ENTER key
. Scroll with the
▲
or
▼
key until the upper display shows the control type (pH,
ORP mV or ORP %) and the lower display shows symmetrical (SY) or asymmetrical (ASY) input type.
Choose ORP without the mV indicator for ORP% readings.
3)
Proceed to 7.6.3
, or return to Measurement mode by pressing the
▲
and
▼
keys simultaneously
(escape).