Instruction Manual MM850470 ed.1a
49
8.2.3
Cal Proc key
[
ª
]
TagSel
¨
(tag files defined by user)
Results
[Found, Left]
Display procedure
Display results
Run procedure
Page 1: ...F lli Casiraghi 409 413 I 20099 Sesto S Giovanni MI Tel 39 022488201 Fax 39 022440286 e mail info eurotron com http www eurotron com M Mi ic cr ro oC Ca al l P P2 20 00 0 2 channels multifunction cali...
Page 2: ...in a retrieval system or transmitted in any form electronic or mechanical without prior written permission of Eurotron Instruments S p A MicroCal P200 Calibrator uses sophisticated analogic and digita...
Page 3: ...old Joint Reference Setting 24 4 1 8 Temperature Scale Setting 26 4 1 9 Temperature Unit Setting 26 4 1 10 Scaling 27 4 2 Source mode 28 4 3 Scaling programming 28 4 5 Data Logging 29 4 6 Transmitter...
Page 4: ...Integrated pressure sensor is installed for gauge absolute and differential measurements An external pressure sensor can be installed for measurements up to 700 bar Report of Calibration Each MicroCa...
Page 5: ...IN EXTREME CIRCUMSTANCES BE LETHAL FURTHERMORE POSSIBLE SERIOUS DAMAGES CAN OCCUR TO THE INSTRUMENT THE USER S SYSTEM AND EQUIPMENT RECHARGEABLE NI MH BATTERIES RECHARGEABLE BATTERIES MUST BE RECYCLED...
Page 6: ...ensor res 0 01mbar 5 0 95 to 2 bar Gauge sensor res 0 01mbar 5A 2 bar Absolute sensor res 0 01mbar 6 0 95 to 7 bar Gauge sensor res 0 1mbar 7 0 95 to 20 bar Gauge sensor res 0 1mbar 7A 20 bar Absolute...
Page 7: ...from 0 95 to 35 bar 500 PSI res 1mbar BB480016 from 0 to 70 bar 1000 PSI res 1mbar BB480017 from 0 to 150 bar 2000 PSI res 1mbar BB480018 from 0 to 350 bar 5000 PSI res 10mbar BB480019 from 0 to 700 b...
Page 8: ...ing shipment If any mechanical damage is noted report the damage to the shipping carrier and then notify Eurotron directly or its nearest agent and retain the damaged packaging for inspection A label...
Page 9: ...01 F 0 01 0 05 C 0 01 0 09 F Pt100 JIS 200 600 C 330 1120 F 0 01 C 0 01 F 0 01 0 05 C 0 01 0 09 F Pt100 SAMA 200 600 C 330 1120 F 0 01 C 0 01 F 0 01 0 05 C 0 01 0 09 F Connections 2 3 and 4 wires Sour...
Page 10: ...tance Sampling time Memory 1500 readings complete with date and time General Calibration self learning technique with automatic procedure Common mode rejection 140 dB at ac operation Normal mode rejec...
Page 11: ...1 2 3 4 5 6 7 8 1 LCD graphic display 2 Keyboard 3 Pressure input 1 8 BSPF 4 ON OFF 5 Channel 1 binding post 6 RS232 connector Female MiniDIN 6pin 7 External pressure module connector Female MiniDIN 8...
Page 12: ...llows you to move across the menu When in output signal mode to change the electrical Channel output value SELECT enter in channels setup MENU Recall setup and configuration menu Return in measure mod...
Page 13: ...V mA or Pt100 the engineering unit and the input mode Pressure value show the pressure internal or external the engineering unit and the input or the output mode Status box Symbols Description Batter...
Page 14: ...symbol indicates that a full charge is required Use only the dedicated battery charger supplied by Eurotron together the instrument CAUTION OLD BATTERIES CAN LEAK AND CAUSE CORROSION NEVER LEAVE RUN...
Page 15: ...en the external loop is not equipped with the loop power supplied The calibrator can be as an example connected directly to a recorder controller etc with input circuits configured for current measure...
Page 16: ...e the battery pack for 8 10 hours by connecting the external battery charger 3 3 Power up Press the ON OFF key to switch the instrument on and off The instrument needs some seconds to check the hardwa...
Page 17: ...ess the key to change the current minutes Use the key to input the minutes Press the key to store the value Press the MENU key to return in measuring mode 3 6 Using the backlight and setting contrast...
Page 18: ...ntrast Press to SET confirm the regulation pressing MENU 3 7 Data Hold Press the HOLD ZERO key to freeze the measure The symbol H will displayed in the status box Press the HOLD ZERO key to release th...
Page 19: ...N it powers up with the last settings Please refers to Chapter 2 4 for electrical connections 4 1 Measure mode 4 1 1 Pressure Press the SELECT key until SELECT Pressure is shown Pi Internal pressure P...
Page 20: ...ring Press the key to confirm the selection and going in measuring mode Press the MENU key to return in measure mode without changes 4 1 3 Pressure Zeroing To zeroing the pressure channel hold down th...
Page 21: ...key to highlight the Current parameter Press the key to select the option Press the MENU key to return in measure mode without changes Current Active loop Passive loop Press to SET Press the key to ch...
Page 22: ...tage MicroCal P200 plus model only can support electrical signal input See table on Technical Specifications chapter Press the SELECT key until SELECT CH is shown SELECT Ch1 Out In Press to SET Press...
Page 23: ...stance thermometers input See table on Technical Specifications chapter Two Three or Four wire RTD connections can be performed for high accuracy measurements Refer to the electrical connection chapte...
Page 24: ...E ELECTRICAL CONNECTION AS SHOWN ON 2 4 CHAPTER Press the key to confirm the selection Press the MENU key to return in measure mode without changes 3 wire Pt100 IEC Pt100 OIML Pt100 USLAB Pt100 US Pt1...
Page 25: ...rature Unit C Rj Ch1 ext 0 000 Rj Ch2 ext 0 000 Scale ITS 90 Press to SET Press the and keys to highlight the Rj Ch1 ext option for channel 1 setting or highlight the Rj Ch2 ext option for channel 2 s...
Page 26: ...ext 0 000 Scale ITS 90 Press to SET Select the Scale option Temperature ITS 90 IPTS68 Press to SET Press the key to highlight the temperature scale used Press the key to confirm the selection and goin...
Page 27: ...uring mode Press the MENU key to return in measure mode without changes 4 1 10 Scaling Press the SELECT key until SELECT CH1 is shown SELECT Ch1 Out In Press to SET Press the key to highlight the In p...
Page 28: ...lectrical signal output See table on Technical Specifications chapter Press the SELECT key until SELECT CH is shown SELECT Ch Out In Press to SET Select the OUT parameter OUT Ch Current Voltage Press...
Page 29: ...rogram the engineering unit Source select the voltage or the current input range 200mV 2V 20V 50mA In Low Change this value with the Zero scale In High Change this value with the Input Full Scale scal...
Page 30: ...Sampling time 1 Press to SET On Off RUN and STOP the function In values Allows to program the low and full scale for the Trx source Out values Allows to program the low and full scale for the mA outp...
Page 31: ...aph channel source In Low Set the graph Low scale In High Set the graph full scale Time Set the sampling time To display the graphic select the graphic display mode by pressing the DISPLAY key 4 7 Swi...
Page 32: ...the following page is shown MENU Configuration Unit Data logger Auxiliary Display ChIn ChP Press to SET Select Auxiliary Auxiliary Trx Graph Switch Test Alarm Enable Press to SET Select Alarm Enable...
Page 33: ...guration Display Power On Set Ch Date Time Rejection 50 Hz Baud rate 9600 Press to SET Select Power On Power On Safe Set Ch Set val Press to SET Safe Channel voltage This way avoids any risk of damage...
Page 34: ...elect the option Press the ESC key to return in measure mode without changes 5 2 Firmware upgrade STFlash The instrument has embedded in its memory a software firmware which is loaded at every turning...
Page 35: ...tion cable BB530203 and turn it on Select the right COM port Serial port Select the same communication velocity expressed in Baud either on PC Baud rate and on the instrument ATTENTION SET THE SAME BA...
Page 36: ...PC with CalpMan 2000 Upload one or more Tag calibration routines test points error band etc from a PC to the internal memory of the instrument in order to automate in field tests and calibrations Sel...
Page 37: ...ns Once installed an icon appears in your Windows programs list Connect the MicroCal to a PC through the serial cable After executing the program you can see a menu like that Firstly set up the serial...
Page 38: ...hannels for recording Available settings are T RH P only if optional environmental module is installed PL PH Pe if internal and or external pressure sensor are installed CH1 CH2 Time Data Acquisition...
Page 39: ...s with related characteristic data To visualize the content of a file you have to select it and click on the 4 th icon View file with notepad After a short time a visualization window of Notepad appea...
Page 40: ...as a traditional calibrator in order to check an instrument under test In this case we conventionally speak of procedure on line as the connection between calibrator instrument and PC is still active...
Page 41: ...ved procedures will remain in the Calibrator memory Delete All Measure before UpLoad Select this option to delete from the calibrator memory ALL OFF LINE data Deleted data include Tag calibration resu...
Page 42: ...plier etc Test and calibration data can be memory stored and downloaded to a PC to document the calibration activity that allows building a quality control chart data bank from a single calibration sh...
Page 43: ...un procedure Press to SET Once loaded data calibration it is possible to check them selecting Display procedure One or more pages appear interchangeable by pressing ENTER SELECT Procedure Tag Tag1 Res...
Page 44: ...ondent values will be in reverse mode Eventually abort the procedure by pressing ESC key A confirmation message will be shown on display Finally press ENTER as required at the end of the procedure The...
Page 45: ...ation speed 9600 19200 38400 57600 115200 User select the user level Advanced Standard Lock standard Factory Date Time DD MM YY hh mm Date set the system date DD MM YY Format Euro USA Time 12H am pm a...
Page 46: ...00 Disp High 0 00 Function Lin Sqr Sqrt Log Decimals X3 Unit 5 charaters for your engineering unit Source 200mV 2V 20V 50mA In Low 0 00 In High 0 00 Disp Low 0 00 Disp High 0 00 Function Lin Sqr Sqrt...
Page 47: ...Min 00 Mode Dot Line Switch Test Pressure Alarm Enable Disable Display ChIn ChP ChOut ChP ChP Graph 8 2 Select Key Press the SELECT key until the channel to be programmed is displayed use the and key...
Page 48: ...0 JIS Current Active loop Passive loop Voltage 200mV 2V 20V Resistance 3wire 500 5k 4wire 500 5k XScale X1 X2 X3 X4 Out Current Active loop Passive loop Voltage 2V 20V 8 2 2 Set Pressure Channel Press...
Page 49: ...Instruction Manual MM850470 ed 1a 49 8 2 3 Cal Proc key TagSel tag files defined by user Results Found Left Display procedure Display results Run procedure...
Page 50: ...e Problem Solutions Over Flashes instead of the related engineering unit and it indicates an overcome of the full scale Insert a value below to the full scale Under Flashes instead of the related engi...
Page 51: ...a temperature from 30 C to 60 C with R H less than 90 If the instrument has been unused for a month check the battery voltage and charge the Ni MH batteries for at least 12 hours 9 3 Spare parts EE620...
Page 52: ...or shows evidence of being damaged as a result of excessive corrosion or current heat moisture or vibration improper specification misapplication misuse or other operating conditions outside of Eurotr...
Page 53: ...L 4 GETTING STARTED 16 Graph 30 H Hold 18 I INTRODUCTORY NOTE 2 K Keyboard 12 L Letter of conformity 52 LinMan 38 M MAINTENANCE 50 Math Functions 27 Measure mode 19 O OPERATIONS 19 Ordering code 6 P P...
Page 54: ...Instruction Manual MM850470 ed 1a 54 W WARRANTY 52 Warranty terms 52...