Instruction Manual MM850341 ed.01a
___________________________________________27____________________________________________
The Raryomatic IR beam profile is a cone descibed by the optical charts shown in section 1. The IR thermometer
must be installed so that the target is bigger than circle described by optics and the optical path is free from
obstruction (steam, dust, etc.).
Install the mounting plate in a suitable place in the most favourable position from a thermal, and mechanical point
of view. Mount on the plate, if it is necessary, the cooling device and/or the purge air ejector. If a cooling jacket
has to be used, use the water outlet on the upper part of the support in order not to create air bubbles in the
jacket.
WARNING
IT ADVISABLE TO USE A PROTECTIVE SCREEN DURING THE THERMOMETER ALIGNMENT OPERATIONS IF THE TARGET HAS A
STRONG RADIATION
.
In order to align the support with the target, look through the support hole (without the thermometer) and find the
best position. It's important for the optic path to be free from any obstacle. In particular, when the surface is
targeted through a hole, the hole diameter has to be big enough according to the distance from the instrument. If
optional laser pinpointing system is installed, switch it on for spot centre viewing.