User Guide Euro Basic
Rev 06 del 11/2011
pag. 9 di 16
-
open the side zip which fastens cover and inferior case
-
slide your hand in between inflatable elements underneath the patient’s buttock.
-
Make sure that there are at least 2.5 - 4cm between patient’s buttock and underlying base.
-
In case the above mentioned distance:
-
Increase if distance is less than 2.5 cm
-
Decrease if distance is more than 4 cm
1
1
4
4
.
.
A
A
L
L
A
A
R
R
M
M
S
S
Inflation pressure of the mattress is constantly monitored by an internal pressure switch.
If alarm activates (EXCLUSION MADE FOR INITIAL INFLATION) we recommend to follow this
procedure in order to identify the cause:
-
check that tubes coming out of the compressor have been plugged in correctly
-
check that the inflatable elements of the mattress and the lateral distribution band are not
broken.
-
Check that connection hoses are not kinked, twisted or entangled to the bed
-
check that inflatable elements are clicked into the clutches of the lateral distribution band
If the alarm continues, please contact customer service without opening the compressor
.
Do not perform any action on the compressor. Maintenance has to be carried out
only by authorised staff.
1
1
5
5
.
.
C
C
P
P
R
R
F
F
U
U
N
N
C
C
T
T
I
I
O
O
N
N
Euro Basic overlay mattress is provided with a fast deflation function (CPR) in order to allow
cardiopulmonary resuscitation to be performed on the patient.
The CPR valves are located in the left-hand side of the head end of the mattress. In normal
conditions, the CPR valve has to remain closed. To close it, click the three valves of the big red
block into plugs present on the red band. Pull gently to verify that the block does not come out.
If necessary, before performing CPR, open CPR valve to release air quickly from the mattress.
To open it, firmly pull out the big red block, pressing on the transparent tongue.