4 Specifications

Operating Range

-125 Pa to +125 Pa

Measuring Unit

Pa (Pascal)

Resolution

16 Bit (0.004 Pa, Calculated)

Accuracy

Dependent on Calibration

Typical Accuracy

±0.2 Pa + 3 % of readings

Drift / Year

< 0.05 Pa/year

Response Time (

τ

)

< 3 ms*

Operating Temperature

-40

°C

to +50

°C

Allowable Overpressure

1 bar

Burst Pressure

5 bar

Sensor Type

Digital Differential Pressure Sensor

Certifications

RoHS

Identification

Unique GS1 GRAI Code

Tubing Connection

Ø4 mm

Dimension

71*55*21 mm

Wire  Length

100 mm

Weight

45 g

Casing

ABS

*Eupry Data Logger has an adjustable measuring interval from 30 seconds to 1 hour
1: Requires a calibration service agreement with Eupry

4.1 Application in Extreme Environment

When handling the Differential Pressure Sensor, it is important not to add any small objects into the two flow channels on
top of the sensor. The sensor is specified for operation in air, N2 or O2. Avoid aggressive or etching substances such as H2O2
or NH3.

Liquid

Avoid any kind of liquid, fatty and oil emulsive vapors inside the flow channel. If exposed to such conditions, it cannot be
guaranteed that the sensors have no drift.

Condensation

Avoid condensation inside the sensor. After condensed water has completely evaporated, the sensor will be fully functional
again.

Dust

Protect the sensor from dust, which could enter into the flow channel.

User Guide: Differential  Pressure  Sensor, P4P1