4 Specifications
Operating Range
-125 Pa to +125 Pa
Measuring Unit
Pa (Pascal)
Resolution
16 Bit (0.004 Pa, Calculated)
Accuracy
Dependent on Calibration
Typical Accuracy
±0.2 Pa + 3 % of readings
Drift / Year
< 0.05 Pa/year
Response Time (
τ
)
< 3 ms*
Operating Temperature
-40
°C
to +50
°C
Allowable Overpressure
1 bar
Burst Pressure
5 bar
Sensor Type
Digital Differential Pressure Sensor
Certifications
RoHS
Identification
Unique GS1 GRAI Code
Tubing Connection
Ø4 mm
Dimension
71*55*21 mm
Wire Length
100 mm
Weight
45 g
Casing
ABS
*Eupry Data Logger has an adjustable measuring interval from 30 seconds to 1 hour
1: Requires a calibration service agreement with Eupry
4.1 Application in Extreme Environment
When handling the Differential Pressure Sensor, it is important not to add any small objects into the two flow channels on
top of the sensor. The sensor is specified for operation in air, N2 or O2. Avoid aggressive or etching substances such as H2O2
or NH3.
Liquid
Avoid any kind of liquid, fatty and oil emulsive vapors inside the flow channel. If exposed to such conditions, it cannot be
guaranteed that the sensors have no drift.
Condensation
Avoid condensation inside the sensor. After condensed water has completely evaporated, the sensor will be fully functional
again.
Dust
Protect the sensor from dust, which could enter into the flow channel.
User Guide: Differential Pressure Sensor, P4P1
Rev. 0.1 -May 2022