Clear light™ Makes Professional IPL Treatment Easier and more Efficient
30
(12) Current shot count
Display the number of current shot.
4.4.5 State display
(12) (13) (14)
(15) (16)
(13) Current shot erase button
(14) Total shot count
(15) Temperature display
(16) Cooling button
ON
OFF
A button for erasing current shot.
Display the number of total shots.
Press the button to show the number of total shots for 2 second.
Display current temperature of filter.
Button for cooling operating. It is automatically operating when power on and ready
mode and the button will be turn to red.
* Cooling automatic system
<Control 1> -5℃ ~ 0℃, <Control 2> 0℃ ~ 15℃
- When the power is on, it is automatically start cooling on and operate <Control 1>
system to cool down to -5℃.
- If it doesn’t have any input for 1 minute, <Control 2> system will be start.
- When it have an operating commend, the cooling will be operate automatically.