U2836 Operation Manual Chapter 3 Introduction to [MEAS DISP]
19
2)
Use above soft keys to set the average times or use numeric keys and
[
ENTER
] input average times directly.
3.3.3
Monitor function
The monitor function can monitor the real voltage across DUT or real current
through DUT. The monitored voltage value is displayed in
Vm
zone on <
MEAS
DISPLAY
> page while the monitored current value is in
Im
zone. The monitor
function can also monitor the test parameter of L, C, R, Z, Y, D, Q,
θ
, X, G and B.
———————————————————————————————————————
Note: The correction function can influence the level monitor function, so when
the correction data changes the level monitor value will change. When the
correction is switched between OPEN or SHOR, the level monitor value will be
influenced as well
———————————————————————————————————————
Operation steps for setting the monitor function
Execute the following operation steps to set the level monitor function as ON or
OFF.
1)
Move the cursor to
MON 1
or
MON 2
zone, the following soft keys will be
displayed.
OFF
V
I
MORE 1/5
2)
Select and press a softkey to set monitor function. If the softkey you need is
not displayed on the page
,
press More to display the following softkeys.
Ls
Lp
Cs
Cp
MORE 2/5
3)
Select and press a softkey to set monitor function. If the softkey you need is
not displayed on the page
,
press More to display the following softkeys.
Rs
Rp
Z
Y
MORE 3/5
4)
Select and press a softkey to set monitor function. If the softkey you need is
not displayed on the page
,
press More to display the following softkeys.
D
Q
θ
r