4-22
Chapter 4 Handler Interface
4.1
Introduction
U2810D provides the Handler interface which can provide the control signals and the
comparison output signals. As described in table 4-1, the control signals include
/TRIG
(External Trigger signal),
/IDX
(Analog measurement completed signal), and
/EOM
(End of
measure and comparison data valid signal). The comparison output signals include
/P1
,
/P2
,
/P3
,
/AUX
and
/NG
. Using these signals, the U2810D can easily be combined with a
component handler and a system controller to fully automate component testing, sorting,
and quality control data processing to increase production efficiency.
Table4-1 Handler interface signal description
4.2
Operation
4.2.1 Signal Definition
The handler interface uses three kinds of signals: comparison output, control input and
control output.
Comparison output signals
:
/P1
,
/P2
,
/P3
,
/NG
,
/AUX. See Figure 4-1 signal’s area example.
Control output signal:
/IDX
(
Analog measurement completed signal
)
/EOM
(
End of measure and comparison data valid signal
)
Control input signal:
/TRIG (External trigger signal)
Signal
Name
Description Specifications
/P1
/P2
/P3
Pass bin signals
/NG
No-Good bin signal
/AUX
Auxiliary bin signal
/IDX
Analog measurement
completed signal
/EOM
End of measurement
Build in pull-up resistor
Open collector output
Negative
true
Opto-isolated
/TRIG
External trigger signal
Pulse width
≥
1
μ
s, uprising edge trigger,
Low level drive current about 5-10mA.